Spacelabs Xhibit Central Station Recalled for Height and Weight Input Error
Spacelabs Healthcare is recalling Xhibit Central Station units because patient height and weight values may be switched, causing incorrect Body Surface Area calculations.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a risk-of-harm product (medical monitoring device) where a calculation error could impact patient care, but no injuries or illnesses have been reported.
Plain-English summary
Spacelabs Healthcare Inc is recalling the Xhibit Central Station, Model 96102, a medical device used to provide clinicians with central monitoring of patient data. The recall involves 1,075 units, with 614 units distributed in the United States and 461 units distributed outside the United States.
The firm received one report that patient height and weight values were switched when input at the central station. This error caused the bedside monitor's Body Surface Area (BSA) calculation to be incorrect. The affected units were distributed across 33 US states, Guam, and numerous countries worldwide.
Healthcare facilities should check their serial numbers against the list provided by the FDA and contact Spacelabs Healthcare for instructions on how to resolve the issue.
The recalled product
- Product
- Spacelabs Healthcare Xhibit Central Station, Model 96102 is used to provide clinicians with central monitoring of patient data for those patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.
- Manufacturer
- Spacelabs Healthcare Inc
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- **SERIAL NUMBERS OF UNITS DISTRIBUTED IN THE US******* SALA0315F01025
- SLA0114F00426
- SLA0114F00434
- SLA0114F00435
- SLA0114F00444
- SLA0114F00445
- SLA0114F00446
- SLA0114F00447
- SLA0114F00448
- SLA0114F00449
- SLA0114F00451
- SLA0114F00452
- SLA0114F00453
- SLA0114F00454
- SLA0114F00457
- SLA0114F00458
- SLA0114F00460
- SLA0114F00461
- SLA0114F00462
- SLA0114F00463
Distribution
Related recalls
Same manufacturer · Spacelabs Healthcare Inc
See all →- ModerateSpacelabs Xhibit Central Station Recalled for DisplayPort Audio/Video Failures
FDA (Devices) · 2026-08-19
- ModerateSpacelabs SL Command Module patient monitor recall due to manufacturing defect
FDA (Devices) · 2026-05-13
- HighMulti-parameter command modules may produce inaccurate cardiac output readings
FDA (Devices) · 2026-04-15
- HighMonitor display cables recalled over inadequate shielding and electromagnetic interference
FDA (Devices) · 2025-10-22
- HighTwelve foot display cables recalled over inadequate shielding and interference
FDA (Devices) · 2025-10-22
Same hazard · calculation error
See all →- HighPinnacle 3 radiation planning software may miscalculate stopping power ratio
FDA (Devices) · 2025-08-13
- HighMedical Imaging Software Bug Affects Diagnostic Calculations on Daylight Saving Time Dates
FDA (Devices) · 2025-03-05
- HighRayStation radiation therapy software recall for density calculation inconsistency
FDA (Devices) · 2025-02-26
- HighElekta MOSAIQ Oncology Information System: Incorrect Image Offset Calculations
FDA (Devices) · 2025-02-19
- HighSpine Surgery Planning Software Calculation Errors in Rod Planning Parameters
FDA (Devices) · 2025-02-12