Multi-parameter command modules may produce inaccurate cardiac output readings
A circuit board issue in Multi-parameter Command Module model 91496 can start cardiac output measurement early in Auto mode, producing errors and inaccurate readings. 1,790 units are affected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall where inaccurate cardiac output readings may lead to delayed care or fluid overload, a risk of harm with no injuries reported in the source.
Plain-English summary
The Multi-parameter Command Module, model 91496, optioned with Masimo or Nellcor, is being recalled. The action covers 1,790 units.
The module has a circuit board issue. When in Auto mode, cardiac output measurement may activate early, resulting in a Bad Curve error with no cardiac output measurement, or an Irregular Curve error where the cardiac output measurement is not automatically excluded. According to the manufacturer this may result in inaccurate readings or the need for additional readings, which may lead to delayed care and/or fluid overload. The FDA classified the action as Class II.
Distribution was worldwide, including nationwide in the United States across states including New York, New Jersey, Alabama, Oklahoma, Iowa, Montana, Wyoming, Minnesota, Texas, Wisconsin and Pennsylvania. The modules are used in hospital monitoring rather than sold to consumers.
Facilities should identify affected units using the serial numbers and UDI codes in the recall notice and follow the instructions issued by the manufacturer.
The recalled product
- Product
- Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
- Manufacturer
- Spacelabs Healthcare, Inc.
- Affected units
- 1,790
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial Numbers: UDI-DI: 10841522106415
- 10841522106422. 1496-223736
- 1496-223737
- 1496-223738
- 1496-224375
- 1496-224376
- 1496-224400
- 1496-224401
- 1496-224402
- 1496-224403
- 1496-224404
- 1496-224405
- 1496-224406
- 1496-223850
- 1496-223851
- 1496-223852
- 1496-224448
- 1496-224956
- 1496-224957
- 1496-224958
Distribution
Related recalls
Same manufacturer · Spacelabs Healthcare, Inc.
See all →- ModerateSpacelabs Xhibit Central Station Recalled for DisplayPort Audio/Video Failures
FDA (Devices) · 2026-08-19
- ModerateSpacelabs SL Command Module patient monitor recall due to manufacturing defect
FDA (Devices) · 2026-05-13
- HighSix foot display cables recalled over inadequate shielding and interference
FDA (Devices) · 2025-10-22
- HighMonitor display cables recalled over inadequate shielding and electromagnetic interference
FDA (Devices) · 2025-10-22
- HighTwelve foot display cables recalled over inadequate shielding and interference
FDA (Devices) · 2025-10-22
Same hazard · inaccurate readings
See all →- HighBlood Oximeter Calibration Error Causes Inaccurate Oxygen Readings
FDA (Devices) · 2024-11-06
- HighSmiths Medical Arterial Kit recalled for pressure monitoring defect
FDA (Devices) · 2024-10-30
- HighGE HealthCare CRITIKON ONE-CUF blood pressure cuffs may give inaccurate readings
FDA (Devices) · 2024-06-12
- SevereCereLink ICP Extension Cable Recalled for Out of Range Readings
FDA (Devices) · 2023-12-20
- HighMasimo SpO2 Ear Sensor May Produce Inaccurate Oxygen Saturation Readings
FDA (Devices) · 2023-06-28
Similar recalls
- SevereTerumo Recalls CDI Blood Parameter Monitoring System for Inaccurate Readings
FDA (Devices) · 2014-05-28
- HighTerumo Recalls CDI 500 Blood Parameter Monitoring System Due to Software Inaccuracies
FDA (Devices) · 2013-07-24
- HighTerumo Recalls CDI 500 Blood Parameter Monitoring System Due to Software Error
FDA (Devices) · 2013-07-24
- HighSpacelabs Ultraview SL Command Modules Recalled for Random Resets
FDA (Devices) · 2015-05-13
- HighMasimo Rad-G Pulse Oximeter May Randomly Power Off, Causing Loss of Monitoring
FDA (Devices) · 2024-04-24
- HighSpacelabs Ultraview SL Command Modules Recalled for Monitoring Suspension
FDA (Devices) · 2015-05-06