Monitor display cables recalled over inadequate shielding and electromagnetic interference
Spacelabs Healthcare is recalling 16,898 DVI display cables because inadequate shielding may allow electromagnetic interference with other electronically sensitive medical devices in use.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: interference with another medical device in use is a stated patient safety risk, and no injuries are reported.
Plain-English summary
DVI display cables, model/catalog number 012-0893-01, 16-inch DVI-I to DVI-I male/male cables used as an external component with patient monitors, are being recalled.
The display cables used for monitors were not in compliance regarding electromagnetic interference from inadequately shielded cables, which may pose a serious risk to patient safety by interfering with the operation of another electronically sensitive medical device in use.
The recall covers 16,898 units, associated with monitor UDI-DIs including (01)10841522131479 for Xprezzon. Distribution was worldwide, including nationwide in the United States. The FDA has classified the action as Class II.
Facilities should identify affected cables and follow Spacelabs' instructions for replacement.
The recalled product
- Product
- Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software Version: N/A Product Description: 16 inch DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with
- Manufacturer
- Spacelabs Healthcare, Inc.
- Hazard
- electromagnetic-interference
- inadequate-shielding
- device-interference
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- (01)10841522131455 (Qube)
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- ModerateOrthofix MINIRAIL external fixation devices recalled for incorrect model labeling
FDA (Devices) · 2026-07-22
- SevereOlympus Ellik Evacuator Recalled for Inadequate Sterilization Risk
FDA (Devices) · 2026-07-22
- HighRapid Refill Continuous Injection System Recalled for Syringe Loosening Risk
FDA (Devices) · 2026-07-22
- SevereABIOMED Impella CP Heart Pump Set Introducer Sheath Leakage Risk
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22