Spacelabs Healthcare, Inc. recalled 3 products on 2025-10-22

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2025-10-22
Critical
0
Units
25,897

Why these products were recalled

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2025-10-22

    Monitor display cables recalled over inadequate shielding and electromagnetic interference

    Spacelabs Healthcare is recalling 16,898 DVI display cables because inadequate shielding may allow electromagnetic interference with other electronically sensitive medical devices in use.

    Product
    Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software Version: N/A Product Description: 16 inch DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with
    Category
    Distribution
    44 states
  • HighFDA (Devices)··2025-10-22

    Twelve foot display cables recalled over inadequate shielding and interference

    Spacelabs Healthcare is recalling 820 twelve-foot DVI display cables because inadequate shielding may allow electromagnetic interference with other electronically sensitive medical devices in use.

    Product
    Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-12 Software Version: N/A Product Description: 12 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with
    Category
    Distribution
    44 states
  • HighFDA (Devices)··2025-10-22

    Six foot display cables recalled over inadequate shielding and interference

    Spacelabs Healthcare is recalling 8,179 six-foot DVI display cables because inadequate shielding may allow electromagnetic interference with other electronically sensitive medical devices in use.

    Product
    Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with
    Category
    Distribution
    44 states