Philips Ingenia Ambition X MRI System Recall for Falling Speaker Hazard
Philips is recalling 11,226 Ingenia Ambition X MRI systems because ceiling speakers may fall due to magnetic attraction.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (falling objects) where injury has not yet been reported, fitting the 'High' severity criteria for theoretical hazards without confirmed illness or injury.
Plain-English summary
Philips Medical Systems Nederlands is recalling 11,226 Ingenia Ambition X Nuclear Magnetic Resonance Imaging (MRI) systems, Model Number 781356. The recall covers all systems distributed nationwide in the United States.
The recall was initiated because there is a potential for the ceiling speakers to fall from their position and be pulled toward the MRI system by the magnetic field. This presents a risk of injury to patients or staff if the speakers detach and strike individuals.
Healthcare facilities should contact Philips Medical Systems Nederlands for instructions on how to address this issue. Consumers and patients should be aware of this potential hazard when using the affected MRI systems.
The recalled product
- Product
- Ingenia Ambition X, Model Number 781356. Nuclear Magnetic Resonance Imaging System
- Manufacturer
- Philips Medical Systems Nederlands
- Category
- Medical Device — MRI System
- Affected units
- 11,226
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All systems
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 64 recalled by Philips Medical Systems Nederlands under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 64 products in this recall →Related recalls
Same manufacturer · Philips Medical Systems Nederlands
See all →- HighPhilips Allura Xper System power supplies may fail unexpectedly
FDA (Devices) · 2021-06-30
- ModeratePhilips Recalls Azurion X-ray Systems Due to Software Defect
FDA (Devices) · 2019-08-07
- HighPhilips MR Systems Recalled for Magnet Energization Device Electrical Shock Risk
FDA (Devices) · 2019-05-08
- ModeratePhilips Medical Systems Recalls Foot Switches for Allura Xper Systems
FDA (Devices) · 2019-04-24
- ModeratePhilips Medical Systems Recalls Foot Switches for Imaging Systems
FDA (Devices) · 2019-04-24
Same hazard · falling object
See all →- SevereLithonia Lighting Recalls RSX Luminaires with Adjustable Tilt Arms
CPSC · 2026-10-22
- HighMaquet PowerLED 300 Concealed Lighting System Recalled for Detachment Risk
FDA (Devices) · 2026-09-30
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- SevereTerex Digger Derrick Trucks Recalled for Weld Defect
NHTSA · 2026-08-17
Similar recalls
- HighPhilips Ingenia Ambition X MRI Systems Alignment Error in MobiView
FDA (Devices) · 2025-05-28
- HighPhilips MRI Systems: Patient support table floor plate installation defect
FDA (Devices) · 2024-04-24
- HighIngenia Ambition X MRI Systems: Mechanical Defect in Hoisting Blocks
FDA (Devices) · 2021-06-16
- HighPhilips Ingenia Ambition X MRI System Tabletop May Fail to Retract
FDA (Devices) · 2021-09-15
- ModeratePhilips Ingenia Ambition MRI Software Recall for Memory Limit Issue
FDA (Devices) · 2026-09-30
- HighMagnetic Resonance System patient support table floor plate may be incorrectly installed
FDA (Devices) · 2024-04-24