The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0891-2019·Announced 2019-03-06

Philips MRI System Recall Due to Ceiling Speaker Magnetic Hazard

Philips is recalling Intera Achieva 1.5T MRI systems because ceiling speakers may be pulled by the magnetic field and fall.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (falling objects due to magnetic attraction) where injury has not yet been reported, fitting the 'High' severity criteria for theoretical hazards without confirmed illness or injury.

Plain-English summary

Philips Medical Systems Nederlands is recalling Intera Achieva 1.5T IT Nova Nuclear Magnetic Resonance Imaging Systems, Model Number 781175. The recall involves 11,226 units that were distributed nationwide in the United States.

The recall is being issued because there is a potential for the ceiling speakers to fall from their position and be pulled to the MRI system by the magnetic field. This hazard could result in injury to patients or staff if the speakers detach and move toward the imaging area.

Healthcare facilities that have these systems should contact Philips Medical Systems Nederlands for instructions on how to address the issue. The source text does not specify a deadline or additional remediation steps beyond the potential hazard description.

The recalled product

Product
Intera Achieva 1.5T IT Nova, Model Number 781175. Nuclear Magnetic Resonance Imaging System
Affected units
11,226

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All systems

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 64 recalled by Philips Medical Systems Nederlands under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 64 products in this recall →

Same manufacturer · Philips Medical Systems Nederlands

See all →

Similar recalls