BD FACSLyric Flow Cytometer Power Supply Failure Risk
BD is recalling FACSLyric flow cytometer power supplies that may fail to power on or stay on, creating risks of fume inhalation and requiring patients to return for repeat biospecimen collection procedures.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall with potential hazards including fume inhalation and respiratory symptoms. No illnesses or injuries have been reported. This meets the rubric criterion for High severity as a risk-of-harm product where injury has not yet been reported.
Plain-English summary
Becton, Dickinson and Company is recalling 139 BD FACSLyric 3L12C flow cytometer power supplies (part number 650781) identified as defective on October 18, 2024.
The affected power supplies may fail to power on or stay on during operation. This can result in fume inhalation, shortness of breath, and coughing among laboratory staff. Additionally, affected instruments may require patients to return for repeat biospecimen collection procedures, which carry associated risks including bruising, pain, and bleeding. Patient specimens may be lost and test result reporting may be delayed.
The power supplies have been distributed nationwide in the United States and to customers in China, Costa Rica, Singapore, and Thailand. BD identified the issue through customer complaints and an increase in field failures and replacements.
Customers are instructed to immediately inspect their inventory for the specific catalog and power supply serial numbers identified in BD's notification letter (sent November 20, 2024). Follow the Actions for Clinical Users guidance provided by BD while continuing normal instrument operation.
The recalled product
- Product
- FACSLyric 3L12C Instrument RUO, Catalog Number 663518, UDI 00382906635181
- Manufacturer
- Becton, Dickinson and Company, BD Bio Sciences
- Affected units
- 139
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Power supply part number 650781 with serial numbers beginning with V2309
- V2318
Distribution
Part of a larger recall action
This product is one of 5 recalled by Becton, Dickinson and Company, BD Bio Sciences under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · Becton, Dickinson and Company, BD Bio Sciences
See all →- ModerateBD Trucount control kits carry incorrect standard deviation values for beads
FDA (Devices) · 2025-08-13
- ModerateBD Trucount 664343 control kits carry incorrect standard deviation values
FDA (Devices) · 2025-08-13
- HighBD FACS 7-Color Setup Beads Recalled Due to Spectral Overlap Errors
FDA (Devices) · 2025-07-09
- ModerateBD FACS setup beads have errors in spectral overlap factors table
FDA (Devices) · 2025-07-09
- HighBD FACSLyric 2L4C Flow Cytometer Power Supply Failure Recall
FDA (Devices) · 2025-01-01
Same hazard · electrical failure
See all →- High
- SevereJaguar Land Rover Recalls Vehicles for DC-DC Converter Defect
NHTSA · 2026-09-17
- High
- HighHalcore Recalls 2025-2026 Horton Ambulances Due to Touchscreen Failure
NHTSA · 2026-08-14
- HighMarion Recalls 2025 Pumper Fire Engines Due to Display Unit Failure
NHTSA · 2026-08-10
Similar recalls
- HighBD FACSLyric Flow Cytometer Recall for Potential False High Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recalled for Potential False High Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recall for Potential False High Absolute Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recalled for Potential False Absolute Counts
FDA (Devices) · 2019-10-16
- SevereBD FACSLyric Flow Cytometers Recalled for Potential False High Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometers Recalled for Potential False High Absolute Counts
FDA (Devices) · 2019-10-16