BD FACS 7-Color Setup Beads Recalled Due to Spectral Overlap Errors
Becton, Dickinson and Company is recalling BD FACS 7-Color Setup Beads (REF 666289, Lots 85637 and 85638) because errors in the spectral overlap factors table may prevent accurate setup of flow cytometry instruments.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall involving a diagnostic device where errors could affect instrument setup and sample analysis reliability. Although no illnesses or injuries are explicitly reported, the risk-of-harm from inaccurate flow cytometry calibration in a diagnostic context qualifies as High severity.
Plain-English summary
Becton, Dickinson and Company, BD Bio Sciences, is recalling BD FACS 7-Color Setup Beads (REF 666289, Lot numbers 85637 and 85638) used in flow cytometry for fluorescence compensation. The recall involves 278 units distributed across multiple U.S. states, Puerto Rico, and numerous international locations including Canada, Mexico, Australia, Japan, and others.
The product is being recalled due to errors in the spectral overlap factors table for the 7-color setup beads. These errors may affect the accurate setup of flow cytometry instruments prior to analyzing samples, potentially compromising the reliability of instrument calibration.
Customers who have received this product should contact Becton, Dickinson and Company for guidance on how to proceed. The product is identified by UDI 382906662897 and lot numbers 85637 and 85638.
The recalled product
- Product
- REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.
- Manufacturer
- Becton, Dickinson and Company, BD Bio Sciences
- Affected units
- 278
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Lot # 85637 & 85638/ UDI: 382906662897
Distribution
Part of a larger recall action
This product is one of 2 recalled by Becton, Dickinson and Company, BD Bio Sciences under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Becton, Dickinson and Company, BD Bio Sciences
See all →- ModerateBD Trucount 664343 control kits carry incorrect standard deviation values
FDA (Devices) · 2025-08-13
- ModerateBD Trucount control kits carry incorrect standard deviation values for beads
FDA (Devices) · 2025-08-13
- ModerateBD FACS setup beads have errors in spectral overlap factors table
FDA (Devices) · 2025-07-09
- HighBD FACSLyric Flow Cytometer Power Supply Failure Risk
FDA (Devices) · 2025-01-01
- HighMedical Device Recall: BD FACSLyric Flow Cytometer Power Supply Failure Risk
FDA (Devices) · 2025-01-01
Same hazard · calibration error
See all →- ModeratePhotera 400 Phototherapy Device Recalled for Incorrect UV Sensor Calibration
FDA (Devices) · 2026-07-22
- ModeratePhotera 600 3-D phototherapy device UV sensor calibration error recall
FDA (Devices) · 2026-07-22
- ModeratePhotera 600 Phototherapy Device UV Sensor Calibration Error
FDA (Devices) · 2026-07-22
- HighPhothera 200 phototherapy calibration certificates had outputs reversed
FDA (Devices) · 2026-06-10
- HighRoche cobas analyzer may repeat previous results for spline calibrated assays
FDA (Devices) · 2026-03-11
Similar recalls
- LowBD FACS 7-color Setup Beads Recalled for Incorrect CD Version
FDA (Devices) · 2012-09-26
- HighBD FACSLyric Flow Cytometer Recalled for Potential False High Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recall for Potential Inaccurate Cell Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recall for Potential False High Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recalled for Potential False Absolute Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometers Recalled for Potential False High Absolute Counts
FDA (Devices) · 2019-10-16