BD FACSLyric Flow Cytometers Recalled for Potential False High Absolute Counts
Becton, Dickinson and Company is recalling eight BD FACSLyric Flow Cytometers because they may produce falsely high absolute counts when used with BD Trucount Tubes.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of harm to patients through inaccurate diagnostic results that could lead to incorrect clinical management, without reported injuries or fatalities.
Plain-English summary
Becton, Dickinson and Company, BD Biosciences is recalling eight units of the BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument (REF: 662877). The recall applies specifically to these instruments when they are used with BD Trucount Tubes. The affected units are identified by serial numbers Z662877000001 through Z662877000008.
The recall is due to a potential issue where the flow cytometers may experience excessive electronic abort counts when used with specific tubes. This issue may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. These inaccurate results may impact clinical decisions, including the initiation of prophylactic therapy for opportunistic infections in HIV patients, care management for immunosuppressed patients, laboratory developed tests, specimen engraftment potency, and the use of leucoreduced blood products for transfusion.
The affected products were distributed in the United States and in numerous foreign countries. Users of these specific instruments should contact Becton, Dickinson and Company for further instructions regarding the recall and potential replacement or repair.
The recalled product
- Product
- BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD Trucount Tubes
- Manufacturer
- Becton, Dickinson and Company, BD Biosciences
- Affected units
- 8
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (8)
- BD FACSLyric Flow Cytometer serial numbers: Z662877000001
- Z662877000002
- Z662877000003
- Z662877000004
- Z662877000005
- Z662877000006
- Z662877000007
- Z662877000008.
Distribution
Part of a larger recall action
This product is one of 9 recalled by Becton, Dickinson and Company, BD Biosciences under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 9 products in this recall →Related recalls
Same manufacturer · Becton, Dickinson and Company, BD Biosciences
See all →- HighBD Multitest 6-Color TBNK Reagent Contamination May Cause Erroneous Results
FDA (Devices) · 2024-07-24
- HighMedical diagnostic reagent recall: incorrect cell staining fluorescence
FDA (Devices) · 2024-01-17
- HighBD Trucount Tubes for Blood Testing Recall Due to Label Detachment
FDA (Devices) · 2022-11-09
- HighBD Trucount Tubes Recalled Due to Partial Label Detachment Causing Operational Failures
FDA (Devices) · 2022-11-09
- ModerateBD Multitest 6-Color TBNK tubes recalled for partial label detachment
FDA (Devices) · 2022-11-09
Same hazard · inaccurate results
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
- HighHardyDisk AST Cefiderocol 30µg for In Vitro Testing Recalled Due to Incorrect Cartridges
FDA (Devices) · 2026-08-12
- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
Similar recalls
- HighBD Trucount Tubes Recalled Due to Partial Label Detachment Causing Operational Failures
FDA (Devices) · 2022-11-09
- ModerateBD Multi-Check CD4 Low Control Recalled for Inaccurate Flow Cytometry Results
FDA (Devices) · 2020-12-16
- ModerateBD Trucount Tubes Recalled for Potential Sealing Defects
FDA (Devices) · 2016-07-27
- ModerateBD Multitest 6-Color TBNK tubes recalled for partial label detachment
FDA (Devices) · 2022-11-09
- ModerateBD Trucount Absolute Counting Tubes Recalled for Lot Mixing
FDA (Devices) · 2016-05-04
- ModerateBD Vacutainer Urine Collection Cups Recalled for Potential RBC Count Elevation
FDA (Devices) · 2018-11-28