BD Trucount Tubes for Blood Testing Recall Due to Label Detachment
Becton, Dickinson is recalling BD Trucount Tubes due to partial label detachment that can cause tubes to stick together or create automation errors. The defect stems from an adhesive supplier change and affects 52,570 units.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall of a diagnostic medical device. Although no illnesses or injuries have been reported, partial label detachment poses a theoretical risk of test failures and automation errors that could affect patient diagnoses through incorrect test results.
Plain-English summary
Becton, Dickinson and Company, BD Biosciences is recalling BD Trucount Tubes (Catalog Number 340334), which are used to determine absolute counts of white blood cells in blood samples. The tubes are designed for use with BD Tritest reagents and compatible flow cytometers.
The recall affects approximately 52,570 tubes distributed in the United States and internationally, including Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay.
The issue is partial label detachment on the tubes. This defect can cause tubes to become stuck together in the pouch or to cause automation errors when used with the automated loader for the wash assistant. Becton, Dickinson's investigation confirmed the problem resulted from a change in adhesive supplied by a second-tier supplier.
Facilities and individuals using these tubes should discontinue use of the affected lot numbers and contact Becton, Dickinson for return and replacement instructions.
The recalled product
- Product
- BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the…
- Manufacturer
- Becton, Dickinson and Company, BD Biosciences
Distribution
Part of a larger recall action
This product is one of 3 recalled by Becton, Dickinson and Company, BD Biosciences under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Becton, Dickinson and Company, BD Biosciences
See all →- HighBD Multitest 6-Color TBNK Reagent Contamination May Cause Erroneous Results
FDA (Devices) · 2024-07-24
- HighMedical diagnostic reagent recall: incorrect cell staining fluorescence
FDA (Devices) · 2024-01-17
- ModerateBD Multitest 6-Color TBNK tubes recalled for partial label detachment
FDA (Devices) · 2022-11-09
- HighBD Trucount Tubes Recalled Due to Partial Label Detachment Causing Operational Failures
FDA (Devices) · 2022-11-09
- HighBD Anti-Kappa PE Flow Cytometry Reagent May Show Inaccurate Results
FDA (Devices) · 2022-06-08
Same hazard · label detachment
See all →- HighMaserati Recalls 2024 Granturismo for Detaching Tire Label
NHTSA · 2025-04-14
- High2023 Kia Sorento Recalled for Detachable Air Bag Warning Labels
NHTSA · 2023-08-15
- High2021 BMW Motorcycles Recalled for Detachable Tire Pressure Labels
NHTSA · 2020-08-20
Similar recalls
- ModerateBD Trucount Tubes Recalled for Potential Sealing Defects
FDA (Devices) · 2016-07-27
- ModerateBD Trucount Absolute Counting Tubes Recalled for Lot Mixing
FDA (Devices) · 2016-05-04
- HighBD FACSLyric Flow Cytometers Recalled for Potential False High Absolute Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recalled for Potential False Absolute Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recall for Potential Inaccurate Cell Counts
FDA (Devices) · 2019-10-16
- HighBD FACSLyric Flow Cytometer Recall for Potential False High Absolute Counts
FDA (Devices) · 2019-10-16