BD FACSLyric Flow Cytometer Recall for Potential False High Absolute Counts
Becton, Dickinson and Company is recalling 236 BD FACSLyric Flow Cytometers because they may produce falsely high absolute counts when used with BD Trucount Tubes.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the defect poses a risk of harm through incorrect clinical decisions (such as unnecessary therapy or transfusion) without reported injuries, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Becton, Dickinson and Company, BD Biosciences is recalling 236 units of the BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument (REF: 659180 and 65918001). The recall applies specifically to instruments used with BD Trucount Tubes. The affected units were distributed across 26 U.S. states and 44 foreign countries.
The recall is due to a potential defect where the flow cytometers may experience excessive electronic abort counts when used with the specific tubes. This issue can affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. The source text does not report any illnesses or injuries associated with this defect.
The inaccurate counts may impact clinical decisions, including the initiation of prophylactic therapy for opportunistic infections in HIV patients, care management for immune-suppressed patients, laboratory developed tests, specimen engraftment potency, and the use of leucoreduced blood products for transfusion. Users should contact Becton, Dickinson and Company for instructions on how to proceed with the recall.
The recalled product
- Product
- BD FACSLyric Flow Cytometer,3-Laser 10 Color Instrument, REF: 659180 and 65918001, when using BD Trucount Tubes
- Manufacturer
- Becton, Dickinson and Company, BD Biosciences
- Affected units
- 236
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- BD FACSLyric Flow Cytometer serial numbers: 9180000057
- 9180000219
- 9180000283
- 0000001R6591800
- R658180000282
- R6591800000082
- R6591800000083
- R659180000045
- R659180000051
- R659180000061
- R659180000063
- R659180000064
- R659180000068
- R659180000071
- R659180000075
- R659180000089
- R659180000099
- R659180000137
- R659180000167
- R659180000196
Distribution
Part of a larger recall action
This product is one of 9 recalled by Becton, Dickinson and Company, BD Biosciences under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 9 products in this recall →Related recalls
Same manufacturer · Becton, Dickinson and Company, BD Biosciences
See all →- HighBD Multitest 6-Color TBNK Reagent Contamination May Cause Erroneous Results
FDA (Devices) · 2024-07-24
- HighMedical diagnostic reagent recall: incorrect cell staining fluorescence
FDA (Devices) · 2024-01-17
- ModerateBD Multitest 6-Color TBNK tubes recalled for partial label detachment
FDA (Devices) · 2022-11-09
- HighBD Trucount Tubes Recalled Due to Partial Label Detachment Causing Operational Failures
FDA (Devices) · 2022-11-09
- HighBD Trucount Tubes for Blood Testing Recall Due to Label Detachment
FDA (Devices) · 2022-11-09
Same hazard · inaccurate results
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
- HighHardyDisk AST Cefiderocol 30µg for In Vitro Testing Recalled Due to Incorrect Cartridges
FDA (Devices) · 2026-08-12
- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
Similar recalls
- ModerateBD Multi-Check CD4 Low Control Recalled for Inaccurate Flow Cytometry Results
FDA (Devices) · 2020-12-16
- HighBD Trucount Tubes Recalled Due to Partial Label Detachment Causing Operational Failures
FDA (Devices) · 2022-11-09
- ModerateBD Trucount Tubes Recalled for Potential Sealing Defects
FDA (Devices) · 2016-07-27
- ModerateBD Multitest 6-Color TBNK tubes recalled for partial label detachment
FDA (Devices) · 2022-11-09
- ModerateBD Trucount Absolute Counting Tubes Recalled for Lot Mixing
FDA (Devices) · 2016-05-04
- HighBD FACSLyric Flow Cytometer Power Supply Failure Risk
FDA (Devices) · 2025-01-01