The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0368-2021·Announced 2020-11-11

Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Concerns

Zimmer Biomet is recalling specific acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II and involves a risk-of-harm product (medical device) where adverse tissue reactions are a potential but not yet reported outcome, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

Zimmer Biomet, Inc. is recalling 412 units of ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene components. The specific product number is 00-8065-648-26, with an internal diameter of 26 mm and an external diameter of 51 mm. The recall covers all lots and the products were distributed nationwide in the United States.

The recall was initiated because the hydrogen peroxide sterilant used in the manufacturing process was not evaluated to confirm the biocompatibility of the residual sterilant. As a result, the manufacturer cannot rule out the potential for adverse tissue reactions in patients who have received these components.

This recall is classified as Class II by the U.S. Food and Drug Administration. Healthcare providers and patients who have received these specific hip prosthesis components should contact Zimmer Biomet for further instructions regarding the recall and potential replacement or monitoring.

The recalled product

Product
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-26
Affected units
412

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All lots

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 75 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 75 products in this recall →

Similar recalls