31 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
31
Critical
0
Severe
6
Most recent
2020-11-11

Of the 31 tissue reaction recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 6 (19%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all tissue reaction recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–25 of 31

  • SevereFDA (Devices)·Z-0344-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Prosthesis Cups Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 acetabular hip prosthesis cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 55 mm Item Number: 00-8065-552-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0338-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0335-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 51 mm Item Number: 00-8065-548-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0345-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 57 mm Item Number: 00-8065-554-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0330-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cup components because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 49 mm Item Number: 00-8065-546-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0375-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 55 mm Number: 00-8065-652-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0359-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cup components because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 45 mm Number: 00-8065-642-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0368-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Concerns

    Zimmer Biomet is recalling specific acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0328-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 47 mm Item Number: 00-8065-544-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0394-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55 mm Number: 00-8065-952-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0355-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cup hip prosthesis components because the hydrogen peroxide sterilant used was not evaluated for biocompatibility.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 61 mm Number: 00-8065-558-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0371-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 53 mm Number: 00-8065-650-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0376-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Components

    Zimmer Biomet is recalling 412 acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 55 mm Number: 00-8065-652-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0370-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling specific ZCA All Poly Acetabular Cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0347-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 57 mm Number: 00-8065-554-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0323-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm Item Number: 00-8065-540-22 Hip prosthesis component
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0341-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cup components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 55 mm Item Number: 00-8065-552-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0332-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32mm, Neutral, O.D. Cup with Spacers, 49 mm Item Number: 00-8065-546-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0383-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0360-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0389-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 61 mm Number: 00-8065-658-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0380-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Components

    Zimmer Biomet is recalling 412 acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 57 mm Number: 00-8065-654-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0391-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 49 mm Number: 00-8065-946-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0684-2017·2016-12-07

    MicroPort Dynasty Liners Recalled Due to Metal Debris Tissue Reaction Risk

    MicroPort Orthopedics is recalling 6,250 Dynasty Liners used in hip replacements due to an increasing trend of suspected tissue reactions to metal debris.

    Product
    Dynasty Liners: REF DLCOGA28, REF DLCOGB32, REF DLCOGC36, REF DLCOGD38, REF DLCOGE40, REF DLCOGF42, REF DLCOGG44, REF DLCOGH48, REF DLCOGJ52, REF DLCOGK56, For use with the conserve metal head
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0688-2017·2016-12-07

    MicroPort Orthopedics Recalls Conserve Cups Due to Metal Debris Tissue Reaction

    MicroPort Orthopedics is recalling Conserve Cups because revision rate trends indicate a specific hazard of tissue reaction to metal debris.

    Product
    Conserve Cups: REF 38BF3642, REF 38BF3642SP, REF 38BF3844, REF 38BF3844SP, REF 38BF4046, REF 38BF4046SP, REF 38BF4248, REF 38BF4248SP, REF 38BF4450, REF 38BF4450SP, REF 38BF4652, REF 38BF4652SP, REF 38BF4854, REF 38BF5056, REF 38BF5258, REF 38BF5258SP, REF 38BF5460, REF…
    Category
    Medical Device
    Distribution
    Distributed nationwide