The Recall Desk
HighFDA (Devices)·Z-0323-2021·Announced 2020-11-11

Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

Zimmer Biomet is recalling 412 hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device where a potential for adverse tissue reactions exists but no specific injuries or illnesses have been reported, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

Zimmer Biomet, Inc. is recalling 412 units of the ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm (Item Number: 00-8065-540-22). These are hip prosthesis components distributed nationwide in the United States.

The recall was initiated because the hydrogen peroxide sterilant used in the manufacturing process was not evaluated to confirm the biocompatibility of the residual sterilant. As a result, the manufacturer cannot rule out the potential for adverse tissue reactions in patients.

This recall affects all lots of the specified item. Consumers and healthcare providers should contact Zimmer Biomet for further instructions regarding the return or replacement of these devices.

The recalled product

Product
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm Item Number: 00-8065-540-22 Hip prosthesis component
Affected units
412

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All lots

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 75 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 75 products in this recall →