HighFDA (Devices)·Z-0339-2021·Announced 2020-11-11

Zimmer Biomet Recalls Acetabular Hip Prosthesis Cups Due to Sterilant Concerns

Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II and involves a risk-of-harm product (medical implant) where the hazard is a potential adverse tissue reaction, but no specific injuries or illnesses have been reported in the source text.

Plain-English summary

Zimmer Biomet, Inc. is recalling 412 units of the ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 53 mm (Item Number: 00-8065-550-28). The recall covers all lots of this specific product, which was distributed nationwide in the United States.

The recall was initiated because the hydrogen peroxide sterilant used in the acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. As a result, the company cannot rule out the potential for adverse tissue reactions in patients who have received these implants.

This recall is classified as Class II by the U.S. Food and Drug Administration. Healthcare providers and patients who have received these specific acetabular cups should contact Zimmer Biomet or their healthcare provider for further instructions regarding the recall and potential next steps.

The recalled product

Product
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-28
Affected units
412

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All lots

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 75 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.