The Recall Desk

FDA (Devices) recall action 86521

Zimmer Biomet, Inc. recalled 75 products on 2020-11-11

The FDA (Devices) publishes a recall like this one as 75 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 75 pages.

Products
75
Announced
2020-11-11
Critical
0
Units
29,252

Why these products were recalled

Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 75

  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 55 mm Number: 00-8065-652-28
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Concerns

    Zimmer Biomet is recalling specific ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-32
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 49 mm Item Number: 00-8065-546-28
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 57 mm Number: 00-8065-654-22
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 59 mm Number: 00-8065-956-32
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Prosthesis Cups Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-28
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-32
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Prosthesis Components Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 61 mm Number: 00-8065-558-32
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 55 mm Item Number: 00-8065-552-28
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-26
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cup components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 55 mm Item Number: 00-8065-552-22
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-22
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 55 mm Number: 00-8065-652-22
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Concerns

    Zimmer Biomet is recalling specific acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-26
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Components

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-26
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-28
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Prosthesis Components Due to Sterilant Concerns

    Zimmer Biomet is recalling specific acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-28
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 61 mm Number: 00-8065-558-26
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cup hip prosthesis components because the hydrogen peroxide sterilant used was not evaluated for biocompatibility.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 61 mm Number: 00-8065-558-28
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 45mm Item Number: 00-8065-542-26 Hip prosthesis component
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 57 mm Item Number: 00-8065-554-22
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-28
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 61 mm Number: 00-8065-958-32
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 61 mm Number: 00-8065-558-22
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 49 mm Number: 00-8065-946-32
    Category
    Distribution
    0 states