Ultrasonic Gastrovideoscope Model TGF-UC180J Recalled for Inadequate Channel Drying
Aizu Olympus is recalling ultrasonic gastrovideoscopes due to an unvalidated channel drying process that left some scopes with water in the channels after repair. Water remaining in the device poses a risk of contamination and infection.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall for a medical device with a theoretical infection risk due to inadequate drying process validation. No illnesses or injuries have been reported. Per the severity rubric, risk-of-harm products where injury has not yet been reported receive a score of 3 (High).
Plain-English summary
Aizu Olympus Co., Ltd. is recalling the ultrasonic gastrovideoscope Model TGF-UC180J. The recall was initiated because the channel air drying process used after repairs was not validated. A small percentage of devices returned to customers after repair contained water in the channels.
Water remaining in the endoscope channels has the potential for contamination that could lead to infection. This device is used for real-time ultrasound imaging, ultrasound-guided needle aspiration, and other endoscopic procedures in the upper gastrointestinal tract and surrounding organs.
Five units of this model were distributed nationwide in the United States. Customers who have received this device should contact Aizu Olympus Co., Ltd. for further information regarding this recall.
The recalled product
- Product
- Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.
- Manufacturer
- Aizu Olympus Co., Ltd.
- Category
- Medical Device — Endoscope
- Affected units
- 5
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Model Number (UDI-DI): TGF-UC180J (04953170328480).
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 22 recalled by Aizu Olympus Co., Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 22 products in this recall →Related recalls
Same manufacturer · Aizu Olympus Co., Ltd.
See all →- LowOlympus OER-Elite devices recalled with a notice on warnings and maintenance
FDA (Devices) · 2025-12-10
- LowOlympus OER-Mini devices recalled with a notice on warnings and maintenance
FDA (Devices) · 2025-12-10
- LowOlympus OER-Pro devices recalled with a notice on warnings and maintenance
FDA (Devices) · 2025-12-10
- HighGastrointestinal Videoscope Recall Due to Improper Repair and Infection Risk
FDA (Devices) · 2023-12-06
- HighColonoscope Repair Defect May Cause Device Malfunction and Contamination Risk
FDA (Devices) · 2023-12-06
Same hazard · water contamination
See all →- HighVFAUOSIT Kitchen Faucets Recalled for Lead Leaching Into Water
CPSC · 2025-05-29
- HighBASDEHEN Kitchen Faucets Recalled Due to Dangerous Lead Exposure
CPSC · 2025-05-29
- HighBronchoscopes Recalled Due to Inadequate Air Drying Process
FDA (Devices) · 2023-11-29
- HighTracheal Fiberscope Airway Management Devices Recalled for Inadequate Channel Drying
FDA (Devices) · 2023-11-29
- HighEndoscopic surgical instruments recalled for unvalidated drying process
FDA (Devices) · 2023-11-29
Similar recalls
- ModerateOlympus Ultrasound Gastrovideoscope Recalled Due to Adhesive Preparation Error
FDA (Devices) · 2019-03-20
- HighUltrasonic Gastrovideoscope Recalled Due to Reprocessing Incompatibility and Infection Control Risk
FDA (Devices) · 2022-05-11
- ModerateOlympus Recalls Gastrovideoscope Due to Insufficient Cleaning Accessory Information
FDA (Devices) · 2015-10-07
- ModerateOlympus GIF-H180J Gastrointestinal Videoscope Recalled Due to Adhesive Error
FDA (Devices) · 2019-03-20
- HighAir/Water Valve Used with Olympus Endoscopes May Malfunction
FDA (Devices) · 2023-08-30
- HighOlympus Ultrasound Endoscope Recall for Updated Cleaning Instructions
FDA (Devices) · 2020-12-09