The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0327-2024·Announced 2023-11-29

Bronchoscopes Recalled Due to Inadequate Air Drying Process

Aizu Olympus recalls bronchoscopes due to unvalidated air drying process. Some returned units after repair contain water in channels, creating infection risk.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: This is a risk-of-harm medical device recall with no reported illnesses or injuries. The hazard—water contamination in invasive airway instruments—creates a realistic infection risk, meeting the rubric criterion for High severity: 'risk-of-harm products where injury has not yet been reported.'

Plain-English summary

Aizu Olympus Co., Ltd. is recalling bronchoscopes used for endoscopy and endoscopic surgery within the airways. The recall affects 1,123 units with worldwide and US nationwide distribution across multiple model numbers.

The air drying process used after repair was not validated. A small percentage of bronchoscopes returned to customers after repair contained water remaining in the channels. Water remaining in the instrument channels has the potential for contamination that could lead to the risk of infection.

This recall addresses the deficiency in the drying process validation that allowed water to remain in equipment after repair.

The recalled product

Product
Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160…
Affected units
1,123

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (18)

  • Model Number (UDI-DI): BF-1T150 (04953170308185)
  • BF-1T60 (04953170308161)
  • BF-1TH190 (04953170335181)
  • BF-1TQ170 (04953170342943)
  • BF-3C160 (04953170340031)
  • BF-3C40 (04953170308130)
  • BF-H190 (04953170335174)
  • BF-MP190F (04953170395581)
  • BF-MP60 (04953170308277)
  • BF-P150 (04953170308178)
  • BF-P190 (04953170342110)
  • BF-P60 (04953170308123)
  • BF-Q170 (04953170342912)
  • BF-Q190 (04953170335198)
  • BF-XP190 (04953170342134)
  • BF-XP60 (04953170308154)
  • BF-XT160 (04953170340147)
  • BF-XT190 (04953170402470).

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 22 recalled by Aizu Olympus Co., Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 22 products in this recall →

Similar recalls