The Recall Desk

FDA (Devices) recall action 93222

Aizu Olympus Co., Ltd. recalled 22 products on 2023-11-29

The FDA (Devices) publishes a recall like this one as 22 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 22 pages.

Products
22
Announced
2023-11-29
Critical
0
Units
15,839

Why these products were recalled

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–22 of 22

  • HighFDA (Devices)··2023-11-29

    Endoscopic surgical instruments recalled for unvalidated drying process

    Aizu Olympus is recalling cystonephroscope endoscopes due to an unvalidated air drying process that allowed water to remain in instrument channels. Retained moisture poses a risk of contamination and infection.

    Product
    Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Ultrasonic Gastrovideoscope Model TGF-UC180J Recalled for Inadequate Channel Drying

    Aizu Olympus is recalling ultrasonic gastrovideoscopes due to an unvalidated channel drying process that left some scopes with water in the channels after repair. Water remaining in the device poses a risk of contamination and infection.

    Product
    Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Olympus endoscopes recalled for potential water contamination in channels

    Aizu Olympus Co., Ltd. is recalling duodenovideoscope models due to an unvalidated channel air drying process. Water remaining in the channels after repair poses a potential contamination risk that could lead to infection.

    Product
    Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Tracheal Fiberscope Airway Management Devices Recalled for Inadequate Channel Drying

    Aizu Olympus is recalling 65 tracheal fiberscope units worldwide due to an unvalidated air drying process. Moisture remaining in device channels after repair could allow contamination and pose an infection risk.

    Product
    Tracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-TP.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Tracheal Videoscope Recall Due to Unvalidated Drying Process

    Aizu Olympus is recalling 61 tracheal videoscopes (Model LF-V) due to an unvalidated air drying process that left water in channels after repair. Water in the endoscope channels may become contaminated and pose an infection risk.

    Product
    Tracheal videoscope- For airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Number: LF-V.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Gastrointestinal videoscopes recalled due to unvalidated channel drying process

    Aizu Olympus is recalling 3,627 gastrointestinal videoscopes because the channel air drying process was not validated. A small percentage of devices returned after repair had wet channels, creating potential for contamination and risk of infection.

    Product
    Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q16
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Mobile Airway Scope Recalled Due to Unvalidated Drying Process

    Aizu Olympus has recalled 33 Mobile Airway Scope units due to an unvalidated air drying process. Some repaired units had wet channels, creating a potential risk of contamination and infection.

    Product
    Mobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Bronchoscopes Recalled Due to Inadequate Air Drying Process

    Aizu Olympus recalls bronchoscopes due to unvalidated air drying process. Some returned units after repair contain water in channels, creating infection risk.

    Product
    Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160, BF-XT
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Colonoscopes recalled over unvalidated channel drying process

    Aizu Olympus colonoscopes are being recalled because the air drying process for the instrument channel was not properly validated. Some units returned after repair had moisture in the channel, creating a risk of contamination and infection.

    Product
    Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-H190DI, PCF-H190DL, PCF-H190I, PCF-H190L, PCF-H190TI, PCF-H190TL, PCF-HQ190I, PCF-HQ190L, PCF-PH190I
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Choleodochoscope Models Recalled Due to Unvalidated Air Drying Process

    Aizu Olympus choleodochoscopes with unvalidated air drying processes may retain water in the channel after repair, risking microbial contamination and patient infection. Affected models: CHF-BP30, CHF-CB30L, CHF-P60.

    Product
    Choleodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Ultrasonic Bronchoscopes Recalled Due to Inadequate Drying Validation

    Aizu Olympus is recalling 823 ultrasonic bronchoscopes (models BF-UC180F, BF-UC190F) because the channel air drying process was not validated, and some repaired units retained water that could cause contamination and infection.

    Product
    Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Sigmoidovideoscope models recalled due to improper channel drying process

    Aizu Olympus sigmoidovideoscope (model PCF-S) units are being recalled due to an unvalidated channel drying process. Some repaired units returned with water in channels, risking contamination and infection.

    Product
    Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Gastroscope Devices Recalled for Inadequate Channel Drying Process

    Aizu Olympus recalls 680 gastroscope-EUS devices due to unvalidated channel drying. Water remaining in the channels poses contamination and infection risks.

    Product
    Gastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, GF-UE160-AL5.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Rhinolaryngoscope Models Recalled for Unvalidated Air Drying Process

    Aizu Olympus rhinolaryngoscopes (models ENF-T3, ENF-VT2, ENF-VT3) are recalled. The air-drying process for the instrument channel was not validated, and water remaining in some devices after repair poses contamination and infection risk.

    Product
    Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-VT3.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Ureteralscopes recalled due to unvalidated drying process and contamination risk

    Aizu Olympus is recalling 1577 ureteralscope units due to an unvalidated channel air drying process that left some units with residual water after repair, creating a contamination and infection risk.

    Product
    Ureteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Transnasal Esophagovideoscope Recalled Due to Unvalidated Drying Process

    Aizu Olympus is recalling the Transnasal Esophagovideoscope (Model PEF-V) because the channel air drying process was not validated, resulting in some units with wet channels that could cause infection.

    Product
    Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Hysterovideoscope Models Recalled Due to Improper Channel Drying Risk

    Aizu Olympus is recalling Hysterovideoscope Model HYF-V units (27 devices) because the channel air drying process was not validated. Water remaining in device channels could pose an infection risk.

    Product
    Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Hysterofiberscope Model HYF-XP Channel Drying Process Flaw Poses Infection Risk

    Aizu Olympus is recalling HYF-XP hysterofiberscopes because the channel air drying process was not validated. Some repaired units have wet channels that could harbor contamination and pose infection risk.

    Product
    Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Small intestinal videoscope recalled due to inadequate channel drying process

    Aizu Olympus is recalling small intestinal videoscopes (Model SIF-Q180) because the channel air drying process was not properly validated. Water remaining in the channels could cause contamination and infection risk.

    Product
    Small intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertion or the lower digestive tract (including the anus, rectum, sigmoid colon, colon, ileocecal valve, a
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Colonoscopes Recalled for Unvalidated Channel Drying Process and Infection Risk

    Aizu Olympus is recalling certain colonoscope models because the channel air drying process was not validated. A small percentage of repaired units had water remaining in the channel, which could lead to infection risk.

    Product
    Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Hysteroscope Model HYF-1T recalled due to moisture in device channels

    Aizu Olympus is recalling Hysteroscope Model HYF-1T devices because the channel air drying process after repair was not validated. Some returned units had residual moisture that could allow contamination and infection.

    Product
    Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-11-29

    Uretero-reno fiberscopes recalled due to inadequate channel drying

    Aizu Olympus is recalling uretero-reno fiberscopes because the channel air drying process was not validated. A small percentage of repaired devices had residual water in the channels, creating a risk of contamination and infection.

    Product
    Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
    Category
    Distribution
    0 states