Sigmoidovideoscope models recalled due to improper channel drying process
Aizu Olympus sigmoidovideoscope (model PCF-S) units are being recalled due to an unvalidated channel drying process. Some repaired units returned with water in channels, risking contamination and infection.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II medical device recall with no reported illnesses or injuries. The hazard poses a risk of infection from contamination in surgical endoscopes, but no incidents have occurred. Per the severity rubric, this qualifies as 'High' for a risk-of-harm product without reported injury.
Plain-English summary
Aizu Olympus Co., Ltd. is recalling sigmoidovideoscope units (model PCF-S, UDI-DI 04953170339912). Sigmoidovideoscopes are medical devices used for endoscopy and endoscopic surgery in the lower digestive tract, including the anus, rectum, and sigmoid colon. Eight units distributed nationwide and worldwide are affected by this recall.
The channel air drying process used on these devices was not properly validated. Following servicing and repair, a small percentage of scopes returned to customers had water remaining in their channels. Water in the channels creates the potential for contamination, which could lead to infection.
Healthcare facilities and medical professionals using affected sigmoidovideoscopes should immediately discontinue use and contact Aizu Olympus Co., Ltd. for further guidance. Devices should not be used until they have been properly serviced to address the wet-channel issue.
The recalled product
- Product
- Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.
- Manufacturer
- Aizu Olympus Co., Ltd.
- Category
- Medical Device — Endoscopy
- Affected units
- 8
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Model Number (UDI-DI): PCF-S (04953170339912).
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 22 recalled by Aizu Olympus Co., Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 22 products in this recall →Related recalls
Same manufacturer · Aizu Olympus Co., Ltd.
See all →- LowOlympus OER-Elite devices recalled with a notice on warnings and maintenance
FDA (Devices) · 2025-12-10
- LowOlympus OER-Pro devices recalled with a notice on warnings and maintenance
FDA (Devices) · 2025-12-10
- LowOlympus OER-Mini devices recalled with a notice on warnings and maintenance
FDA (Devices) · 2025-12-10
- HighGastrointestinal Videoscope Recall Due to Improper Repair and Infection Risk
FDA (Devices) · 2023-12-06
- HighColonoscope Repair Defect May Cause Device Malfunction and Contamination Risk
FDA (Devices) · 2023-12-06
Same hazard · water contamination
See all →- HighVFAUOSIT Kitchen Faucets Recalled for Lead Leaching Into Water
CPSC · 2025-05-29
- HighBASDEHEN Kitchen Faucets Recalled Due to Dangerous Lead Exposure
CPSC · 2025-05-29
- HighEndoscopic surgical instruments recalled for unvalidated drying process
FDA (Devices) · 2023-11-29
- HighTracheal Fiberscope Airway Management Devices Recalled for Inadequate Channel Drying
FDA (Devices) · 2023-11-29
- HighGastrointestinal videoscopes recalled due to unvalidated channel drying process
FDA (Devices) · 2023-11-29
Similar recalls
- HighEVIS EXERA III Gastrointestinal Videoscope Recalled Due to Sterilization Failures
FDA (Devices) · 2023-11-15
- HighOlympus colonoscope repair defect may impair imaging and suction function
FDA (Devices) · 2023-12-06
- ModerateOlympus PCF-Q180AC Video Colonoscope Recalled Due to Adhesive Error
FDA (Devices) · 2019-03-20
- HighColonoscope Repair Defect May Cause Device Malfunction and Contamination Risk
FDA (Devices) · 2023-12-06
- HighGastrointestinal Videoscope Recall Due to Improper Repair and Infection Risk
FDA (Devices) · 2023-12-06
- HighImproperly Repaired Gastrointestinal Videoscope May Expose Patients to Contamination
FDA (Devices) · 2023-12-06