Philips Mobile Detector Holder Recalled Due to Potential Detector Drop Hazard
Philips is recalling 412 Mobile Detector Holders for Stationary X-ray Systems because the lock mechanism may fail, causing the detector to drop.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (falling heavy equipment) where injury has not yet been reported, fitting the 'High' severity criteria for theoretical hazards without confirmed illness or injury.
Plain-English summary
Philips Electronics North America Corporation is recalling 412 units of the Mobile Detector Holder for Stationary X-ray Systems (RF System). These devices are used to position wireless portable detectors in the room for X-ray exposures and support the detector with or without a grid. The affected units were manufactured and distributed between March 2015 and March 2017.
The recall is due to a defect in the lock mechanism of the movable clamp. If the mechanism is not adjusted properly or secured, the portable detector can drop when the movable clamp is at the bottom and additional force is applied. This hazard may occur when the detector is rotated clockwise from a landscape to a portrait position.
Healthcare facilities and users of these devices should stop using the affected Mobile Detector Holders immediately. Philips is the recalling firm and is responsible for managing the recall. Users should contact Philips for instructions on how to resolve the issue or obtain a replacement.
The recalled product
- Product
- Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.
- Manufacturer
- Philips Electronics North America Corporation
- Affected units
- 412
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Philips Electronics North America Corporation
See all →- HighPhilips Xper Flex Cardio Physiomonitoring System Recalled for Display Errors
FDA (Devices) · 2018-12-26
- ModeratePhilips IntelliVue MX40 Patient Monitor Speaker Volume Recall
FDA (Devices) · 2018-12-19
- HighPhilips Medical Monitor Ceiling Suspension Recall for Detachment Risk
FDA (Devices) · 2018-10-03
- HighPhilips Allura Xper FD2O Ceiling Monitor Suspension Recall
FDA (Devices) · 2018-10-03
- HighPhilips AlIura Xper F010 F Monitor Ceiling Suspension Recall
FDA (Devices) · 2018-10-03
Same hazard · falling object
See all →- SevereLithonia Lighting Recalls RSX Luminaires with Adjustable Tilt Arms
CPSC · 2026-10-22
- HighMaquet PowerLED 300 Concealed Lighting System Recalled for Detachment Risk
FDA (Devices) · 2026-09-30
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- SevereTerex Digger Derrick Trucks Recalled for Weld Defect
NHTSA · 2026-08-17
Similar recalls
- HighPhilips Digital Diagnost Mobile Detector Holder Recall for Insecure Fit
FDA (Devices) · 2012-10-03
- HighPhilips DXR System DuraDiagnost Rel. 4 X-Ray System Collimator Fall Risk
FDA (Devices) · 2025-04-23
- HighPhilips DXR System X-Ray Equipment Collimator Fall Hazard Recall
FDA (Devices) · 2025-04-23
- HighPhilips Digital Diagnost Radiographic System Recall Due to Footboard Defect
FDA (Devices) · 2014-04-30
- SeverePhilips DuraDiagnost X-Ray System Recalled for Installation Defects
FDA (Devices) · 2015-03-18
- SeverePhilips OmniDiagnost-Eleva X-ray System Recalled Due to Table Rotation Hazard
FDA (Devices) · 2018-07-04