Leica Biosystems Cryostat Model CM3050 S lacks warning label for flammable sprays
Leica cryostat units lack a warning about flammable freezing sprays that can ignite in the chamber. The company discovered through post-market surveillance that users were applying these flammable sprays without being warned of the fire hazard.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall of a risk-of-harm product with no reported injuries or hospitalizations. The hazard involves potential ignition of flammable sprays within the device, but since no injury has been reported and the risk is theoretical, the severity is 3 per the rubric.
Plain-English summary
Leica Biosystems is recalling Cryostat Model Number CM3050 S units due to missing safety labeling. The cryostat is used for freezing and sectioning tissue samples in laboratory and medical settings. Post-market surveillance identified that customers were using flammable freezing sprays inside the cryostat chamber without being warned of the flammability hazard.
Flammable freezing sprays can ignite within the cryostat, potentially causing serious injuries. While the flammability of these sprays is marked on the spray containers themselves, the cryostat labeling did not include a specific warning about this risk prior to 2019.
A total of 187 units have been distributed nationwide in the United States. The affected product includes multiple material numbers and serial numbers as specified by the FDA.
Customers using affected units should consult updated safety information and instructions from Leica Biosystems before continuing to use flammable sprays with the cryostat.
The recalled product
- Product
- Leica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 14903050S03; for freezing and sectioning tissue samples
- Manufacturer
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Affected units
- 187
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- a) Material Number 14047033518
- UDI/DI 04049188065599
- Serial Numbers: 5013012008
- 3151052002
- 3559082003
- b) Material Number 14903050S01
- UDI/DI 04049188003270
- Serial Numbers: 3641
- 4119012005
- 5212102008
- 4002102004
- 3944092004
- 3137042002
- 3683122003
- 5119
- 4125022005
- 4171032005
- 3954092004
- 2636102000
- 3847052004
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by LEICA BIOSYSTEMS NUSSLOCH GMBH under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · LEICA BIOSYSTEMS NUSSLOCH GMBH
See all →- HighLeica Biosystems Cryostat CM1950 Firmware Update May Cause Tissue Loss
FDA (Devices) · 2025-03-05
- HighCryostat Model CM1950 Lacks Warning on Flammable Spray Use
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1100 Recalled for Fire Risk from Flammable Freezing Sprays
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1850 Lacks Fire-Risk Warning for Flammable Sprays
FDA (Devices) · 2024-10-23
- HighLeica Biosystems Cryostat lacks warning for flammable freezing spray use
FDA (Devices) · 2024-10-23
Same hazard · fire hazard
See all →- Critical
- HighLomi roll-on waxing kit power cords can overheat and short circuit
CPSC · 2026-06-25
- HighAmana room air conditioners may stay energized during a ground fault
CPSC · 2026-06-25
- HighMito Red Light Super Mobile batteries may catch fire
FDA (Devices) · 2026-06-24
- High
Similar recalls
- HighLeica Cryostats Recalled for Missing Flammable Spray Warning
FDA (Devices) · 2019-06-12
- HighLeica Biosystems Cryostat CM1950 Firmware Update May Cause Tissue Loss
FDA (Devices) · 2025-03-05
- SevereLeica HistoCore SPECTRA ST Slide Stainer Recalled for Electrical Safety Hazard
FDA (Devices) · 2018-10-10
- ModerateMRI Liquid Cooling Cabinet may have incorrect electrical shock warning label
FDA (Devices) · 2024-01-31
- ModerateMRI Cooling Cabinet Has Incorrect Electrical Warning Label
FDA (Devices) · 2024-01-31
- ModerateLeica CD15 Antibody Recalled for Decreasing Staining Intensity
FDA (Devices) · 2015-04-01