Cryostat Model CM1950 Lacks Warning on Flammable Spray Use
Leica Biosystems is recalling the Cryostat Model CM1950 due to missing warnings about using flammable freezing sprays in the device. Flammable sprays can ignite inside the chamber, potentially causing serious injuries.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall for a medical device with potential fire and serious injury hazard from flammable sprays. No reported illnesses or injuries stated in source. Fits rubric criterion for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Leica Biosystems Nussloch GmbH is recalling 30 units of the Cryostat Model CM1950 distributed nationwide in the United States. The device is used in laboratory and medical settings for freezing and sectioning tissue samples.
The recall addresses a labeling deficiency: the cryostat's labeling did not include specific warnings about the use of flammable freezing sprays inside the device chamber. Post-market surveillance revealed that some customers were using flammable freezing sprays for this purpose. Because these sprays are flammable, they can potentially ignite inside the chamber, creating a fire hazard that could cause serious injuries to laboratory personnel.
The affected units are identified by their material numbers and serial numbers as documented by the manufacturer. Customers with these units should contact Leica Biosystems for updated labeling and guidance.
The recalled product
- Product
- Leica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c) 1491950C2US, d) 1491950C3US; for freezing and sectioning tissue samples
- Manufacturer
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Affected units
- 30
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- a) Material Number 14047742467
- UDI/DI 04049188053992
- Serial Numbers: 492102008
- b) Material Number 1491950C1US
- UDI/DI 04049188002501
- Serial Numbers: 432082008
- 288032008
- 581122008
- 435082008
- 380
- 215012008
- 427082008
- 218012008
- 287032008
- 216012008
- 291042008
- c) Material Number 1491950C2US
- UDI/DI 04049188002488
- Serial Numbers: 524112008
- 451082008
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by LEICA BIOSYSTEMS NUSSLOCH GMBH under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · LEICA BIOSYSTEMS NUSSLOCH GMBH
See all →- HighLeica Biosystems Cryostat CM1950 Firmware Update May Cause Tissue Loss
FDA (Devices) · 2025-03-05
- HighLeica Biosystems Cryostat Model CM3050 S lacks warning label for flammable sprays
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1850 Lacks Fire-Risk Warning for Flammable Sprays
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1900 Lacks Warning on Flammable Sprays
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1100 Recalled for Fire Risk from Flammable Freezing Sprays
FDA (Devices) · 2024-10-23
Same hazard · fire
See all →- SevereDREO Recalls DR-HSH015 Space Heaters Due to Fire and Burn Hazards
CPSC · 2026-10-29
- SeverePamag Pieeoo Mattresses Recalled for Fire Hazard Risk
CPSC · 2026-10-29
- SevereHAIARA Power Banks Recalled Due to Fire and Burn Hazards
CPSC · 2026-10-29
- SevereWhatever Works Recalls Microwave Grills Due to Fire Hazard
CPSC · 2026-10-29
- CriticalNTI and HTP Gas Boilers Recalled for Fire and Carbon Monoxide Risks
CPSC · 2026-10-29
Similar recalls
- HighLeica Cryostats Recalled for Missing Flammable Spray Warning
FDA (Devices) · 2019-06-12
- HighLeica Biosystems Cryostat CM1950 Firmware Update May Cause Tissue Loss
FDA (Devices) · 2025-03-05
- SevereLeica HistoCore SPECTRA ST Slide Stainer Recalled for Electrical Safety Hazard
FDA (Devices) · 2018-10-10
- ModerateMRI Liquid Cooling Cabinet may have incorrect electrical shock warning label
FDA (Devices) · 2024-01-31
- HighLeica Surgical Microscopes Recalled for Potential Overheating and Smoke
FDA (Devices) · 2016-04-27
- HighParaffin Embedding Station Recalled for Internal Fire and Toxic Smoke Risk
FDA (Devices) · 2024-08-14