Leica Biosystems Cryostat lacks warning for flammable freezing spray use
Leica Biosystems Cryostat Model CM1510 S is recalled due to missing warning labels about flammable freezing sprays, which can ignite and cause serious injury. The device labeling lacked fire hazard warnings prior to 2019.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II medical device recall for a risk-of-harm hazard involving potential fire and serious injury from flammable freezing sprays. The source text does not report any actual illnesses, injuries, or deaths, so per the rubric this is High (3) rather than Severe, consistent with the criterion for risk-of-harm products without reported harm.
Plain-English summary
Leica Biosystems is recalling the Cryostat Model Number CM1510 S (Material Numbers 1491510C002 and 1491510C003), which is used to freeze and section tissue samples in clinical and research laboratories. A total of 13 units were distributed nationwide in the US.
Post-market surveillance identified that some customers are using flammable freezing sprays inside the cryostat chamber. While flammable freezing sprays are labeled with flammability warnings, the cryostat device labeling did not include any specific warning about the use of flammable freezing sprays prior to 2019. Because all flammable freezing sprays can potentially ignite inside the chamber, they present a risk of serious injury.
The recalled product
- Product
- Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples
- Manufacturer
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Affected units
- 13
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (16)
- a) Material Number 1491510C002
- UDI/DI (01)04049188002945
- Serial Numbers: 5359022006
- 5367012006
- 5913
- 5996
- 5065
- 5625022007
- 5467072006
- 5901
- 5108122004
- 5330
- 5269
- 5667
- b) Material Number 1491510C003
- UDI/DI (01)04049188002891. Serial Numbers: 5537
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by LEICA BIOSYSTEMS NUSSLOCH GMBH under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · LEICA BIOSYSTEMS NUSSLOCH GMBH
See all →- HighLeica Biosystems Cryostat CM1950 Firmware Update May Cause Tissue Loss
FDA (Devices) · 2025-03-05
- HighLeica Biosystems Cryostat Model CM3050 S lacks warning label for flammable sprays
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1850 Lacks Fire-Risk Warning for Flammable Sprays
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1900 Lacks Warning on Flammable Sprays
FDA (Devices) · 2024-10-23
- HighLeica Cryostat Model CM1100 Recalled for Fire Risk from Flammable Freezing Sprays
FDA (Devices) · 2024-10-23
Same hazard · fire
See all →- SeverePrepared Hero Recalls Fire Spray Canisters Due to Rupture Risk
CPSC · 2026-09-10
- SevereSweet Cubby Recalls Children's Loungewear Sets Due to Burn Hazard
CPSC · 2026-09-10
- Severe
- Severe
- SevereFord Recalls 2023-2027 F-150 Trucks for Fuel Tank Securing Defect
NHTSA · 2026-09-09