Citrine QTRAP Mass Spectrometer Recalled for Flammable Solvent Fire Hazard
AB Sciex is recalling ten Citrine QTRAP Mass Spectrometer units due to potential flame and electrode ejection from flammable solvent leaks combined with electrical discharge.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The source text does not report any actual illnesses, injuries, or hospitalizations. The hazard is described as potential and rare, making it theoretical rather than realized. This qualifies as a risk-of-harm product where injury has not yet been reported, placing it in the High severity category per the rubric.
Plain-English summary
AB Sciex, LLC is recalling ten units of the Citrine QTRAP Mass Spectrometer (Part Number 5063685) due to a potential fire and equipment damage hazard. Under rare circumstances, a leak of highly flammable solvent at the ion source probe, combined with an electrical discharge, may result in a flame on the probe and possible ejection of the electrode assembly.
This hazard poses risks to users, patients, and the surrounding environment. If left unattended, the flame could cause personal injury, damage to the equipment, or damage to nearby property. Additionally, this malfunction causes delays in obtaining patient test results.
The recalled units have the following serial numbers: DD250051901, DD250172009, DD250182009, DD230001804, DD250111910, DD250132004, DD250091909, DD250101910, DD250192010, and CN10011707PT. These units were distributed nationwide in the United States.
Users of affected Citrine QTRAP Mass Spectrometer units should immediately discontinue use and contact AB Sciex, LLC for guidance on replacement, repair, or return options. Laboratories should verify their equipment serial numbers against the recall list to determine if their instruments are affected.
The recalled product
- Product
- Citrine QTRAP Mass Spectrometer, Part Number 5063685
- Manufacturer
- AB Sciex, LLC
- Affected units
- 10
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (11)
- UDI-DI: 00842365100540
- Serial Numbers: DD250051901
- DD250172009
- DD250182009
- DD230001804
- DD250111910
- DD250132004
- DD250091909
- DD250101910
- DD250192010
- CN10011707PT
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 4 recalled by AB Sciex, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · AB Sciex, LLC
See all →- HighMass Spectrometer Recalled for Risk of Flame and Electrode Ejection
FDA (Devices) · 2023-10-11
- HighAPI3200MD Mass Spectrometer Recalled for Potential Fire Risk
FDA (Devices) · 2023-10-11
- HighAB Sciex 4500MD Triple Quad Mass Spectrometer Recall Due to Fire Risk
FDA (Devices) · 2023-10-11
Same hazard · fire
See all →- SevereDREO Recalls DR-HSH015 Space Heaters Due to Fire and Burn Hazards
CPSC · 2026-10-29
- SeverePamag Pieeoo Mattresses Recalled for Fire Hazard Risk
CPSC · 2026-10-29
- SevereHAIARA Power Banks Recalled Due to Fire and Burn Hazards
CPSC · 2026-10-29
- SevereWhatever Works Recalls Microwave Grills Due to Fire Hazard
CPSC · 2026-10-29
- CriticalNTI and HTP Gas Boilers Recalled for Fire and Carbon Monoxide Risks
CPSC · 2026-10-29
Similar recalls
- HighAB Sciex Mass Spectrometers Recalled for Incorrect Quantitative Results
FDA (Devices) · 2016-05-25
- HighAb Sciex QTRAP 4500MD Mass Spectrometer Software Recall for Incorrect Results
FDA (Devices) · 2016-08-24
- ModerateAb Sciex 3200MD Mass Spectrometer Software Recall for Incorrect Results
FDA (Devices) · 2016-08-24
- SeverePhilips UNIQ X-ray Systems Recalled for Potential Fire Hazard
FDA (Devices) · 2017-04-26
- HighAb Sciex Mass Spectrometer Software Recall for Incorrect Quantitative Results
FDA (Devices) · 2016-08-24
- HighPhilips iCT SP UPS Accessory Devices Recalled for Battery Acid Leakage Risk
FDA (Devices) · 2015-06-17