FDA (Devices) recall action 92810

AB Sciex, LLC recalled 4 products on 2023-10-11

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-10-11
Critical
0
Units
65

Why these products were recalled

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2023-10-11

    Mass Spectrometer Recalled for Risk of Flame and Electrode Ejection

    AB Sciex is recalling 14 Citrine Triple Quad Mass Spectrometers (Part Number 5063684) due to rare risk of solvent leak and electrical discharge potentially causing flame and electrode ejection at the ion source probe.

    Product
    Citrine Triple Quad Mass Spectrometer, Part Number 5063684
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-10-11

    AB Sciex 4500MD Triple Quad Mass Spectrometer Recall Due to Fire Risk

    AB Sciex has recalled 38 units of the 4500MD Triple Quad Mass Spectrometer nationwide due to a rare risk of flame and electrode ejection from solvent leaks combined with electrical discharge.

    Product
    4500MD Triple Quad Mass Spectrometer, Part Number 5032522
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-10-11

    API3200MD Mass Spectrometer Recalled for Potential Fire Risk

    AB Sciex is recalling three API3200MD mass spectrometers due to a rare risk of fire and electrode ejection from a solvent leak combined with electrical discharge. The flame could cause injury to users or property damage.

    Product
    API3200MD Mass Spectrometer, Part Number 4466230
    Category
    Distribution
    0 states