Ab Sciex QTRAP 4500MD Mass Spectrometer Software Recall for Incorrect Results
Ab Sciex is recalling 28 QTRAP 4500MD mass spectrometers because MultiQuant MD software may display incorrect quantitative results when using the Sum Multiple Ions feature.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device that may provide incorrect diagnostic results, which poses a risk of harm to patients, but no specific injuries or illnesses have been reported in the source text.
Plain-English summary
Ab Sciex is recalling 28 units of the QTRAP 4500MD LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. The affected units are equipped with MultiQuant MD software versions 3.0, 3.0.1, and 3.0.2. The recall is due to a software defect where, under certain conditions, users may be presented with incorrect quantitative results when utilizing the Sum Multiple Ions feature.
The affected instruments were distributed worldwide, including in the United States and internationally to Canada, Austria, China, the Netherlands, Norway, India, Korea, Germany, France, Switzerland, Ireland, Italy, Singapore, Russia, the UK, and Malaysia. These mass spectrometers are intended to identify inorganic or organic compounds, such as lead, mercury, and drugs, in human specimens for in vitro diagnostic purposes.
Users of the affected software should contact Ab Sciex for further instructions regarding the recall and potential corrective actions.
The recalled product
- Product
- AB SCIEX QTRAP¿ 4500MD LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human…
- Manufacturer
- Ab Sciex
- Affected units
- 28
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Software Version: MultiQuant MD 3.0
- MultiQuant MD 3.0.1 and MultiQuant MD 3.0.2 software
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 4 recalled by Ab Sciex under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Ab Sciex
See all →- HighCliquid MD Software Version 3.4 May Provide Incorrect Analytical Results
FDA (Devices) · 2021-11-03
- HighAb Sciex API 3200MD Software Recall for Incorrect Quantitative Results
FDA (Devices) · 2016-08-24
- ModerateAb Sciex 3200MD Mass Spectrometer Software Recall for Incorrect Results
FDA (Devices) · 2016-08-24
- HighAb Sciex Mass Spectrometer Software Recall for Incorrect Quantitative Results
FDA (Devices) · 2016-08-24
- HighAB Sciex Mass Spectrometers Recalled for Incorrect Quantitative Results
FDA (Devices) · 2016-05-25
Similar recalls
- HighAb Sciex API 3200MD System Recalled for Incorrect Diagnostic Data
FDA (Devices) · 2014-09-17
- HighAb Sciex Analyst MD Software Recall for Incorrect Diagnostic Results
FDA (Devices) · 2016-05-11
- HighAB Sciex Mass Spectrometers Recalled for Incorrect Quantitative Results
FDA (Devices) · 2016-05-25
- HighAB Sciex 4500MD Triple Quad Mass Spectrometer Recall Due to Fire Risk
FDA (Devices) · 2023-10-11
- HighCitrine QTRAP Mass Spectrometer Recalled for Flammable Solvent Fire Hazard
FDA (Devices) · 2023-10-11
- ModerateAgilent MassHunter Software Recall Due to Sample Data Mismatch
FDA (Devices) · 2019-09-04