  • HighFDA (Devices)·Z-1641-2026·2026-04-01

    Olympus HALO PKS cutting forceps jaws may break from defective welds

    Olympus is recalling HALO PKS Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electro
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1644-2026·2026-04-01

    Olympus PKS cutting forceps jaws may break from defective welds

    Olympus is recalling PKS Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visuali
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1635-2026·2026-04-01

    Medtronic MiniMed 780G software defects may cause insulin delivery inaccuracies

    Medtronic is recalling MiniMed 780G insulin pumps on software 6.60 and 6.61 because three software defects could result in hypoglycemia or hyperglycemia through inaccurate insulin delivery.

    Product
    MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1640-2026·2026-04-01

    Olympus PK cutting forceps jaws may break from defective welds

    Olympus is recalling PK Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alternating current to soft tissues for cutting and coagulation dur
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1596-2026·2026-04-01

    I.T.S. pilon and tibia plates heat more in MRI than stated

    I.T.S. GmbH is recalling Pilon Plates with Angular Stability and Anterolateral Tibia Plates because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 2109
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·26V209000·2026-04-01

    Audi Q3 recalled over remote sunroof closing from out of sight

    Volkswagen Group of America is recalling 6,077 Audi Q3 vehicles because incorrect software may allow remote sunroof closing from a distance where the operator cannot clearly see the sunroof.

    Product
    AUDI — AUDI Q3
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1597-2026·2026-04-01

    I.T.S. olecranon plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Olecranon Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1590-2026·2026-04-01

    GE Imactis CT navigation may show needle trajectory that differs from actual

    GE Medical Systems is recalling the Imactis CT-Navigation System because of a potential discordance between the simulated needle trajectory and tip position and the actual trajectory and tip position.

    Product
    Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1601-2026·2026-04-01

    I.T.S. fibula and tubular plates heat more in MRI than stated

    I.T.S. GmbH is recalling the Fibula Plate PROlock with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibul
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1604-2026·2026-04-01

    I.T.S. distal ulna plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling the Distal Ulna Locking and Ulna Osteotomy Locking Systems because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·26V208000·2026-04-01

    Volvo trucks recalled over turn signals that may not function properly

    Volvo Trucks North America is recalling 340 VNL and VNR vehicles because the turn signals may not function properly, failing to comply with the federal lighting standard.

    Product
    VOLVO — VOLVO VNL (4), VNR (4)
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1634-2026·2026-04-01

    TECNIS Eyhance intraocular lens haptic may stick to the optic

    AMO Puerto Rico Manufacturing is recalling TECNIS Eyhance IOLs with the TECNIS Simplicity Delivery System because a limited quantity may have a haptic that sticks to the optic, preventing consistent unfolding.

    Product
    Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1600-2026·2026-04-01

    I.T.S. pelvic reconstruction screws heat more in MRI than stated

    I.T.S. GmbH is recalling the Pelvic Reconstruction System fixation screw and short spike because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1593-2026·2026-04-01

    I.T.S. clavicle plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Clavicula Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1633-2026·2026-04-01

    Straumann angled custom abutments made from wrong bar material

    Straumann is recalling Custom Abutment AS Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1639-2026·2026-04-01

    Olympus Everest bipolar cutting forceps jaws may break from defective welds

    Olympus is recalling Everest Bipolar Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achi
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1643-2026·2026-04-01

    Olympus PKS cutting forceps 920000PK jaws may break from welds

    Olympus is recalling PKS Cutting Forceps with cord, model 920000PK, because a supplier did not adequately validate the welding process and defective welds can cause the jaw to break.

    Product
    Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1608-2026·2026-04-01

    I.T.S. pelvic reconstruction plates heat more in MRI than stated

    I.T.S. GmbH is recalling the Pelvic Reconstruction System PRS RX and PRS Phoenix because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0405-2026·2026-04-01

    Teva octreotide acetate 20 mg recalled over sterility assurance deficiencies

    Teva is recalling nine lots of Octreotide Acetate for Injectable Suspension 20 mg after quality system deficiencies were identified during an FDA inspection at the contract manufacturer.

    Product
    OCTREOTIDE ACETATE — OCTREOTIDE ACETATE (OCTREOTIDE ACETATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1629-2026·2026-04-01

    Orthalign drill plate sizes 1-5 may allow excessive bone cut

    Orthalign is recalling one lot of the Trimax Pegged Drill Plate sizes 1-5 for the Lantern Knee System because plates manufactured out of specification can result in a bone cut in excess of 0.0800 inches.

    Product
    Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1606-2026·2026-04-01

    I.T.S. headless compression screws heat more in MRI than stated

    I.T.S. GmbH is recalling the HCS Headless Compression Screw System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0590-2026·2026-04-01

    Chilli Spanish Peanuts Recalled for Glass Contamination Risk

    Western Mixers Produce & Nuts, Inc. is recalling Treasured Harvest brand Chilli Spanish Peanuts (25lb bags, Lot #260230) due to glass fragments found in sealed containers.

    Product
    Chilli Spanish Peanuts, Treasured Harvest brand, 25lb clear plastic bag container. UPC:none Item code: 2608-25
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1612-2026·2026-04-01

    Olympus OER-Elite reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Elite Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1599-2026·2026-04-01

    I.T.S. cannulated screw system heats more in MRI than stated

    I.T.S. GmbH is recalling the I.T.S. Screw System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42, 314
    Category
    Medical Device
    Distribution
    Distributed nationwide