I.T.S. distal ulna plates heat more in MRI than instructions state
I.T.S. GmbH is recalling the Distal Ulna Locking and Ulna Osteotomy Locking Systems because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported: implanted hardware may heat more during MRI than its instructions state, and no injuries are stated in the source.
Plain-English summary
I.T.S. GmbH is recalling the I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System, covering distal ulna plates, ulna osteotomy plates and locking cancellous screws. About 120 units were distributed across 30 US states and territories. The FDA has classified this recall as Class II.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use.
Affected product is listed by article number and lot number in the FDA enforcement record.
Facilities can identify affected implants by the article and lot numbers, and may contact I.T.S. GmbH for further information.
The recalled product
- Product
- I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6…
- Manufacturer
- I.T.S. GmbH
- Affected units
- 120
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1. Article Number (Lot Numbers): 21232-3 (26/F65636)
- 21232-4 (26/F61722
- 26/F65637)
- 21232-6 (26/F62564
- 26/F65638)
- 21233-3 (26/F65639)
- 21233-4 (26/F65640)
- 21233-6 (26/F61723
- 26/F65046)
- 21234-4 (26/F65641
- 26/F66504)
- 21235-4 (26/F65642
- 26/F67810). 2. Article Number (Lot Numbers): 21231-6 (643/012301
- 643/022310
- 643/032308
- 643/0630
- 643/0657
- 643/102402
- 643/122301). 3. Article Number (Lot Numbers): 37302-18 (18/636447)
- 37302-20 (18/802189).
Distribution
Part of a larger recall action
This product is one of 17 recalled by I.T.S. GmbH under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 17 products in this recall →Related recalls
Same manufacturer · I.T.S. GmbH
See all →- HighI.T.S. volar radius plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
- HighI.T.S. proximal humeral plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
- HighI.T.S. straight and forearm plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
- HighI.T.S. olecranon plates heat more in MRI than instructions state
FDA (Devices) · 2026-04-01
- HighI.T.S. hand locking plates heat more in MRI than instructions state
FDA (Devices) · 2026-04-01
Same hazard · mri heating
See all →- HighI.T.S. pilon and tibia plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
- HighI.T.S. pelvic reconstruction screws heat more in MRI than stated
FDA (Devices) · 2026-04-01
Similar recalls
- ModerateI.T.S. GmbH Recalls Titanium Rim Plates Due to Incorrect Laser Marking
FDA (Devices) · 2019-06-12
- HighMedartis Recalls APTUS Ulna Shortening 2.5 Plates Due to Breakage
FDA (Devices) · 2014-11-05
- HighElekta Unity Radiation Therapy System Recalled for Heating Issues
FDA (Devices) · 2020-07-08
- HighPivotal Health Solutions Recalls Apollo Laser Desktop Control Units
FDA (Devices) · 2015-03-18
- SevereCook Inc. Recalls Bush Ureteral Illuminating Catheter Sets Due to Overheating Risk
FDA (Devices) · 2017-04-26
- ModerateIntegra bioBLOCK Resorbable Subtalar Implant Recalled for Temperature Exposure
FDA (Devices) · 2019-04-17