The Recall Desk

Hazard

Rf Induced Heating recalls

17 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
17
Critical
0
Severe
0
Most recent
2026-04-01

Of the 17 rf induced heating recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all rf induced heating recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–17 of 17

  • HighFDA (Devices)·Z-1607-2026·2026-04-01

    I.T.S. hand locking plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling the HLS Hand Locking Plates System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1604-2026·2026-04-01

    I.T.S. distal ulna plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling the Distal Ulna Locking and Ulna Osteotomy Locking Systems because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1598-2026·2026-04-01

    I.T.S. straight and forearm plates heat more in MRI than stated

    I.T.S. GmbH is recalling Straight Plates with Angular Stability, including forearm and compression plates, because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1600-2026·2026-04-01

    I.T.S. pelvic reconstruction screws heat more in MRI than stated

    I.T.S. GmbH is recalling the Pelvic Reconstruction System fixation screw and short spike because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1605-2026·2026-04-01

    I.T.S. foot locking plates heat more in MRI than stated

    I.T.S. GmbH is recalling the FLS Foot Locking Plates System and HOL Hallux Osteotomy Locking Plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1594-2026·2026-04-01

    I.T.S. proximal humeral plates heat more in MRI than stated

    I.T.S. GmbH is recalling the Humeral Head Plate with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Brand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal Humeral Plate. Article Numbers: 211341-10, 211341-4, 211341-7, 21134-4, 21134-5, 21135-4, 21136-4, 21136-5, 21136-7, 21137-4, 21137-7, 21138-4, 21139-4.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1593-2026·2026-04-01

    I.T.S. clavicle plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Clavicula Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1599-2026·2026-04-01

    I.T.S. cannulated screw system heats more in MRI than stated

    I.T.S. GmbH is recalling the I.T.S. Screw System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1597-2026·2026-04-01

    I.T.S. olecranon plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Olecranon Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1592-2026·2026-04-01

    I.T.S. volar radius plates heat more in MRI than stated

    I.T.S. GmbH is recalling Volar Radius Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1603-2026·2026-04-01

    I.T.S. femur and tibia plates heat more in MRI than stated

    I.T.S. GmbH is recalling the LRS Locking Reconstruction System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301-5, 21301-7, 21301-9, 21301-13, 21302-5, 21302-7, 21302-9, 21302-13. 2. PTL (Proximal Lateral…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1608-2026·2026-04-01

    I.T.S. pelvic reconstruction plates heat more in MRI than stated

    I.T.S. GmbH is recalling the Pelvic Reconstruction System PRS RX and PRS Phoenix because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1606-2026·2026-04-01

    I.T.S. headless compression screws heat more in MRI than stated

    I.T.S. GmbH is recalling the HCS Headless Compression Screw System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1595-2026·2026-04-01

    I.T.S. calcaneus repair plate heats more in MRI than stated

    I.T.S. GmbH is recalling the FR.O.H. Calcaneus Repair System anterolateral calcaneus plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1596-2026·2026-04-01

    I.T.S. pilon and tibia plates heat more in MRI than stated

    I.T.S. GmbH is recalling Pilon Plates with Angular Stability and Anterolateral Tibia Plates because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1601-2026·2026-04-01

    I.T.S. fibula and tubular plates heat more in MRI than stated

    I.T.S. GmbH is recalling the Fibula Plate PROlock with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1602-2026·2026-04-01

    I.T.S. distal humeral plates heat more in MRI than stated

    I.T.S. GmbH is recalling Distal Humeral Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N…
    Category
    Medical Device
    Distribution
    Distributed nationwide