I.T.S. clavicle plates heat more in MRI than instructions state
I.T.S. GmbH is recalling Clavicula Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported: implanted hardware may heat more during MRI than its instructions state, and no injuries are stated in the source.
Plain-English summary
I.T.S. GmbH is recalling the Claviculaplate with Angular Stability, covering medial anterior clavicle plates, clavicle plates, 3.5 mm clavicle plates and lateral clavicle plates across twelve article numbers. About 322 units were distributed across 30 US states and territories. The FDA has classified this recall as Class II.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use.
Affected product is listed by article number and lot number in the FDA enforcement record.
Facilities can identify affected implants by the article and lot numbers, and may contact I.T.S. GmbH for further information.
The recalled product
- Product
- Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm…
- Manufacturer
- I.T.S. GmbH
- Affected units
- 322
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1. Article Number (Lot Numbers): 21116-8 (679/092230)
- 21117-10 (679/032313
- 679/092231)
- 21118-10 (679/032314
- 679/072047
- 679/092232). 2. Article Number (Lot Numbers): 21121-6 (679/092236
- 679/092329)
- 21121-8 (679/052226). 3. Article Number (Lot Numbers): 21122-10 (00/0425-02
- 00/0825-02
- 679/022443
- 679/062310
- 86/043372)
- 21122-6 (00/0325-01
- 00/0425-01
- 679/042340
- 679/062334
- 679/122269
- 86/043366)
- 21122-8 (00/0325-08
- 679/022243
Distribution
Part of a larger recall action
This product is one of 17 recalled by I.T.S. GmbH under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 17 products in this recall →Related recalls
Same manufacturer · I.T.S. GmbH
See all →- HighI.T.S. volar radius plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
- HighI.T.S. proximal humeral plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
- HighI.T.S. straight and forearm plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
- HighI.T.S. olecranon plates heat more in MRI than instructions state
FDA (Devices) · 2026-04-01
- HighI.T.S. hand locking plates heat more in MRI than instructions state
FDA (Devices) · 2026-04-01
Same hazard · mri heating
See all →- HighI.T.S. distal ulna plates heat more in MRI than instructions state
FDA (Devices) · 2026-04-01
- HighI.T.S. pilon and tibia plates heat more in MRI than stated
FDA (Devices) · 2026-04-01
Similar recalls
- ModerateI.T.S. GmbH Recalls Titanium Rim Plates Due to Incorrect Laser Marking
FDA (Devices) · 2019-06-12
- HighBiomet Recalls A.L.P.S. Clavicle Plating System Due to Drill Fracture Risk
FDA (Devices) · 2021-01-27
- ModerateTrilliant Surgical Gridlock Ankle Plating System Instructions for Use Recalled
FDA (Devices) · 2020-04-08
- HighPhilips MRI Systems Recalled Due to Incorrectly Installed Patient Table Floor Plate
FDA (Devices) · 2024-04-24
- HighOrthopedic Femur Plate Recalled for Incorrect MRI Safety Information
FDA (Devices) · 2021-12-22
- HighMedtronic Atlantis Anterior Cervical Plate Screw Module Recall
FDA (Devices) · 2012-10-17