Dexcom G7 app defect may delay glucose readings and alerts
Dexcom G7 CGM iOS and watchOS app versions 2.9.0 through 2.11.2 are being corrected because a software defect can delay glucose value and alert processing.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II correction where delayed glucose values and alerts could cause a missed hypoglycemic or hyperglycemic event, a risk of harm with no injuries reported.
Plain-English summary
The Dexcom G7 Continuous Glucose Monitoring System watchOS app, model SW13355, software versions 2.9.0 through 2.11.2, is subject to correction, along with the corresponding iOS app SW12300. The action covers 58,582 units.
The issue is a software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app runs in the background, EGV notifications may queue up and then replay sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous readings. In more severe instances the app may perceive signal loss and deploy signal loss mitigations. According to the manufacturer, this delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information. The FDA classified the action as Class II.
Distribution was worldwide. The apps are used by people managing diabetes.
Users should check their app version and follow the update instructions issued by Dexcom. Anyone unsure about a reading should confirm with a fingerstick test in line with their care plan.
The recalled product
- Product
- Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS…
- Manufacturer
- Dexcom, Inc.
- Affected units
- 58,582
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Lot Code: SW13355 versions 2.9.0 through 2.11.2 - UDI-DI: 00386270005168
Distribution
Part of a larger recall action
This product is one of 3 recalled by Dexcom, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Dexcom, Inc.
See all →- SevereDexcom G7 Continuous Glucose Monitoring System sensors recalled
FDA (Devices) · 2026-07-01
- HighDexcom ONE+ iOS CGM App Software Defect Delayed Glucose Alerts
FDA (Devices) · 2026-05-20
- HighDexcom G7 CGM iOS App Software Defect Delays Glucose Alerts
FDA (Devices) · 2026-05-20
- CriticalGlucose monitoring app recalled because it can terminate and miss alerts
FDA (Devices) · 2025-11-05
- CriticalDexcom G7 watchOS app recalled because sensor failure alerts are missing
FDA (Devices) · 2025-09-10
Same hazard · software defect
See all →- SevereMedtronic MiniMed Go App Recall for Insulin Dose Reminder Failure
FDA (Devices) · 2026-10-28
- SeverePhilips Lumify Diagnostic Ultrasound System Recalled for Connection Issues
FDA (Devices) · 2026-10-28
- ModerateIntelerad IntelePACS Software Recalled for Image Display Mismatch
FDA (Devices) · 2026-10-21
- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
Similar recalls
- CriticalDexcom G7 watchOS app recalled because sensor failure alerts are missing
FDA (Devices) · 2025-09-10
- HighDexcom G7 iOS App version 2.7.0 stops updating glucose readings
FDA (Devices) · 2025-06-11
- CriticalDexcom G7 iOS app recalled because sensor failure alerts do not appear
FDA (Devices) · 2025-09-10
- CriticalGlucose monitor apps recalled because sensor failed alerts do not appear
FDA (Devices) · 2025-09-10
- CriticalGlucose monitoring app recalled because it can terminate and miss alerts
FDA (Devices) · 2025-11-05
- HighDexcom G6 Glucose Monitor May Fail to Deliver Alarms Properly
FDA (Devices) · 2025-02-26