Hamilton-C3 Ventilator Recalled Due to Control Board Component Defect
Hamilton Medical AG is recalling two Hamilton-C3 ventilators because an out-of-specification control board component may cause the device to shut down and trigger alarms.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a high-risk medical device (ventilator) where a defect could cause a shutdown, but the source text does not report any injuries, illnesses, or fatalities, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Hamilton Medical AG is recalling two units of the Hamilton-C3 Ventilator. The recall is due to an out-of-specification component on the control board. This defect may cause the ventilator to open the inspiratory channel and the expiratory valve, leading to a device shutdown. When this occurs, the ventilator activates an emergency buzzer and a blinking alarm LED on the front of the device.
The affected devices are identified by serial numbers 3084 and 3088. The Hamilton-C3 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants, and neonates. These devices are used in intensive care wards, intermediate care wards, emergency wards, long-term acute care hospitals, recovery rooms, and during patient transfers.
The recall covers two devices distributed in the United States to California, Colorado, Illinois, Louisiana, Minnesota, New York, Oklahoma, Virginia, and Wisconsin. The devices were also distributed worldwide to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, and Armenia. Healthcare facilities should stop using the affected units and contact the manufacturer for instructions.
The recalled product
- Product
- Hamilton-C3 Ventilator Unit Product Usage: The HAMILTON-C3 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants and neonates depending on its configuration. The device is used in the intensive care ward, intermediate care…
- Manufacturer
- Hamilton Medical AG
- Category
- Medical Device — Ventilator
- Affected units
- 2
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Serial Number: 3084
- 3088
Distribution
Part of a larger recall action
This product is one of 2 recalled by Hamilton Medical AG under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Hamilton Medical AG
See all →- SevereVentilator breathing circuit sets may obstruct exhalation valve
FDA (Devices) · 2026-07-08
- HighIntelliCuff Portable Automatic Cuff Pressure Controller Recall
FDA (Devices) · 2026-07-08
- CriticalHamilton C6 ventilators recalled over circuit board defect interrupting ventilation
FDA (Devices) · 2025-08-27
- SevereHamilton coaxial breathing circuits may have a crack in blue tubing
FDA (Devices) · 2025-08-06
- SevereVentilator HAMILTON-C6 may fail to restart, risking oxygen deprivation
FDA (Devices) · 2024-06-26
Same hazard · device malfunction
See all →- SevereMedtronic MiniMed Flex Insulin Pumps Recalled for Failed Verification Tests
FDA (Devices) · 2026-10-14
- ModerateWilson-Cook Recalls Hemospray Endoscopic Hemostat Due to Adherence Issues
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption
FDA (Devices) · 2026-09-23
Similar recalls
- CriticalHamilton C6 ventilators recalled over circuit board defect interrupting ventilation
FDA (Devices) · 2025-08-27
- HighHamilton-C3 Ventilator Battery Fuel Gauge Recall
FDA (Devices) · 2018-04-18
- SevereHamilton C3 ventilator software anomaly may cause unexpected device stoppage
FDA (Devices) · 2023-08-02
- HighHamilton Medical Recalls HAMILTON-C2 Ventilators Due to Loudspeaker Defects
FDA (Devices) · 2013-11-13
- SevereHAMILTON C1 Ventilator Recalled for Software Defect Causing Unexpected Shutdown
FDA (Devices) · 2023-08-02
- SevereHamilton C2 Ventilator recalled for software anomaly causing loss of function
FDA (Devices) · 2023-08-02