SevereFDA (Devices)·Z-2167-2025·Announced 2025-08-06

Hamilton coaxial breathing circuits may have a crack in blue tubing

Coaxial breathing circuit sets for HAMILTON-C1, T1 and MR1 ventilators are being recalled because the blue tubing that delivers fresh gas may be cracked, which can cause rebreathing of exhaled gases. 2,560 units affected.

What this means for you

Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: FDA Class I recall, which the rubric floors at Severe; no injury is reported in the source, so it does not escalate to Critical.

Plain-English summary

The breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators, UDI-DI 07630002802963, is being recalled. The action covers 2,560 units from lots 199675 and 199676.

According to the manufacturer, the ventilator coaxial breathing circuit may have a crack in the blue tubing, which is used to deliver fresh gas to the patient. The notice states the crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis and, if unrecognised, organ damage. The FDA classified the action as Class I, its most serious classification.

Distribution was nationwide in the United States, in New Hampshire, New York, Connecticut, Pennsylvania, Illinois, Maine, Colorado, North Carolina and New Jersey among other states.

Facilities should identify affected stock using the REF, UDI-DI and lot numbers above, quarantine it, and follow the instructions issued by the manufacturer.

The recalled product

Product
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Manufacturer
Hamilton Medical AG
Hazard
  • cracked-tubing
  • rebreathing
  • hypercapnia
  • ventilator
  • class-i

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • UDI-DI: 07630002802963
  • Lots: 199675
  • 199676

Distribution

Distributed nationwide across the United States.