Innovision Innocor Cardiopulmonary Testing System Recalled for Electric Shock Risk
Innovision A/S is recalling eight Innocor Cardiopulmonary Exercise and Lung Function Testing Systems to replace USB cables that lack galvanic isolation, preventing potential electric shock.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall addresses a risk of electric shock, which is a risk-of-harm product issue where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Innovision A/S is recalling eight Innocor Cardiopulmonary Exercise and Lung Function Testing Systems. These are non-invasive diagnostic devices that measure or calculate hemodynamic and metabolic parameters. The recall involves replacing the data transmission cable (USB-cable) that connects the Innocor device to an external computer.
The recall is necessary because the current USB cable uses grounding protection, which does not provide galvanic isolation. The replacement cable will use opto-couplers to provide galvanic isolation between the Innocor device and the external computer. This change is intended to prevent the risk of electric shock.
The affected units were distributed worldwide, including the United States (Massachusetts and Washington, DC), Canada, Germany, and the United Kingdom. The specific units are identified by Ref. INN00400 and serial numbers 1551369, 1540362, 1540363, and 1536361. Users should contact the manufacturer to arrange for the replacement of the USB cable.
The recalled product
- Product
- Innocor Cardiopulmonary Exercise and Lung Function Testing System Non-invasive diagnostic device that measures or calculates a number of hemodynamic and metabolic parameters.
- Manufacturer
- Innovision A/S
- Hazard
- Affected units
- 8
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Ref. INN00400
- Serial no. 1551369
- 1540362
- 1540363
- and 1536361.
Distribution
Related recalls
Same hazard · electric shock
See all →- High
- High
- Severe
- ModeratePanasonic Electric Toaster Ovens Recalled for Shock and Fire Hazards
CPSC · 2026-07-16
- SevereAltec Recalls 2022-2023 Aerial Work Platforms Due to Missing Check Valve
NHTSA · 2025-12-23
Similar recalls
- ModerateCARDIOHELP-i Heart-Lung Support System not properly tested for leakage current
FDA (Devices) · 2023-12-27
- HighGE Healthcare Recalls CardioLab Recording Systems Due to Electrical Shock Risk
FDA (Devices) · 2019-11-06
- HighEdwards Lifesciences IntraClude Intra-Aortic Occlusion Device Recalled for Potential Damage
FDA (Devices) · 2015-06-17
- SevereEdwards Lifesciences Recalls IntraClude Intra-Aortic Occlusion Devices Due to Balloon Rupture Risk
FDA (Devices) · 2019-06-19
- HighGE Healthcare Recalls Invasive Blood Pressure Care Cables
FDA (Devices) · 2012-07-25
- ModerateInvivo Recalls Xper Flex Cardio PhysioMonitoring Systems Due to Communication Errors
FDA (Devices) · 2017-11-15