Siemens e.cam and Symbia E Gamma Camera Patient Bed Recall
Siemens is recalling e.cam and Symbia E gamma camera systems because patient hair can become caught in the patient bed drive mechanism.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product where injury has not yet been reported, which corresponds to a High severity score of 3 according to the rubric.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 5,451 e.cam and Symbia E gamma camera systems with patient beds. The recall is due to reports that patient hair can become caught in the patient bed up/down drive mechanism.
The affected systems are used to detect or image the distribution of radionuclides in the body or organ using planar imaging, whole body imaging, and tomographic imaging. The recall covers units distributed worldwide, including the United States, Puerto Rico, and numerous international locations.
Siemens is reminding users of the patient positioning instructions and will perform an update on the systems through local Siemens service representatives.
The recalled product
- Product
- e.cam and Symbia E gamma camera systems with patient beds To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV.
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Hazard
- Affected units
- 5,451
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model numbers 4380213
- 4380221
- 5242826
- 5977066
- 5977074
- 5984005
- 5989079
- 5989087
- 5989095
- 5991109
- 5991117
- 5992099
- 7324119
- 7324135
- 7324143
- 7324150
- 7760809
- 7760932
- 7761161
- 7823920
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc.
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · entrapment
See all →- Severe
- Severe
- Severe
- Severe
- Severe
Similar recalls
- HighSiemens e.cam Gamma Camera Systems Recalled for Loose Bolts
FDA (Devices) · 2014-10-29
- ModerateSiemens E.cam and Symbia Systems Recalled for Detector Time Loss
FDA (Devices) · 2020-09-02
- ModerateSiemens Symbia E Camera System Recalled for Loose Rear Casters
FDA (Devices) · 2014-07-23
- HighSiemens e.cam Gamma Camera Recall for Radial Brake Failure
FDA (Devices) · 2015-11-11
- HighSiemens e.cam Systems Recalled for Potential Patient Finger Injury
FDA (Devices) · 2015-01-14
- ModerateSiemens e.cam camera systems recalled for loose rear casters
FDA (Devices) · 2014-07-23