Siemens Symbia Intevo SPECT/CT Systems Recalled for Rubber Isolator Aging
Siemens Medical Solutions is recalling 48 Symbia Intevo SPECT/CT systems because a rubber isolator may degrade over time, potentially leading to drive belt rupture.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a mechanical component degradation that could lead to drive belt rupture, representing a risk-of-harm product where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Siemens Medical Solutions USA, Inc. is performing a field correction on 48 Symbia Intevo series SPECT/CT systems, including the Intevo 16, Intevo 6, Intevo 2, and Intevo Excel models. The recall is due to a rubber isolator within the rotate motor mounting assembly that has unfavorable aging properties. This component may degrade over time, which could impact motor support.
The affected systems were distributed worldwide, including in the United States, Puerto Rico, and numerous international locations. The source text notes that while the isolator degradation does not pose a hazard by itself, it could lead to a rupture of the drive belt. The provided text is truncated and does not specify further consequences or specific instructions for consumers beyond the field correction.
The recalled product
- Product
- The Symbia Intevo series consists of the Intevo 16, Intevo 6 and Intevo 2. This manual also describes the Symbia Intevo Excel. The Symbia Intevo series integrates state-of-the-art SPECT and high quality spiral CT to provide precise attenuation correction and anatomical mapping. T
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Affected units
- 48
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1) Symbia Intevo Excel
- Product Code: 10764801
- Serial Numbers: 1001
- 1002
- 1003
- 1004
- 1005
- 1006
- 1007
- 1008
- 1009
- 1010
- 1011
- 1012
- 1013
- 1014
- 2) Symbia Intevo 2
- Product Code: 10764802
- 2001
- 3) Symbia Intevo 6
Distribution
Part of a larger recall action
This product is one of 2 recalled by Siemens Medical Solutions USA, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc.
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · mechanical failure
See all →- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighForest River Transit Buses Recalled for Wheelchair Lift Control Failure
NHTSA · 2026-09-01
- HighBlue Bird Recalls Vision School Buses Due to Wheelchair Lift Defect
NHTSA · 2026-08-31
- SevereMagic Tilt Watercraft Trailers Recalled Due to Winch Weld Failure
NHTSA · 2026-08-28