DxH 600 hematology analyzers recalled over erroneously high hemoglobin results
UniCel DxH 600 Coulter Cellular Analysis Systems configured with the LED HGB photometer are recalled because they may produce erroneously high hemoglobin results on samples with elevated white blood cell counts.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported; the notice states the analyzer can produce erroneously high patient results and no injuries are reported.
Plain-English summary
The UniCel DxH 600 COULTER Cellular Analysis System configured with the LED HGB Photometer, REF B23858, UDI 15099590192242, is being recalled. Affected serial numbers include AV52143, AZ02628, AZ06676, AZ24836, BA07048, BA08667, BA17727 and BA31855, with the full list in the recall notice. Distribution was nationwide in the United States and internationally.
Hematology analyzers configured with HGB photometers may produce erroneously high hemoglobin results on samples with elevated white blood cell counts. The FDA classified the action as Class II.
Laboratories operating these analyzers should follow the manufacturer's instructions, including any guidance on reviewing results already reported.
The recalled product
- Product
- UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858
- Manufacturer
- Beckman Coulter Inc.
- Hazard
- erroneous-results
- hematology-analyzer
- class-ii
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (11)
- UDI: 15099590192242/ Serial Numbers: AV52143
- AZ02628
- AZ06676
- AZ24836
- BA07048
- BA08667
- BA17727
- BA31855
- BB10648
- BB49881
- BC24713
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- HighMedtronic DLP One-Piece Pediatric Arterial Cannula Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- SevereMedtronic Bio-Medicus Adult Cannula Kit Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- SevereT-Lok Bone Marrow Tray Kits Recalled for Sterility Testing Failure
FDA (Devices) · 2026-07-22
- SevereAbiomed Impella CP Smart Assist Set Heart Pump Introducer Sheath Leakage
FDA (Devices) · 2026-07-22