DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance…
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
Severity not yet assessed
We have not scored this recall against our rubric yet, so we are not showing a severity rating or our usual guidance. The agency’s own notice below is the authoritative source — read it first.
Plain-English summary
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
The recalled product
- Product
- DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
- Manufacturer
- Beckman Coulter, Inc.
- Category
- Medical Device — Devices
- Affected units
- 2,088
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Product/UDI-DI: DxU 840m/15099590748654
- DxU 850m/15099590748647. All Serial Numbers
Distribution
Related recalls
Same manufacturer · Beckman Coulter, Inc.
See all →- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- HighBeckman Coulter DxC 500 AU Analyzer Software Defect Recall
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- High