The Recall Desk

FDA (Devices) recall action 97464

Beckman Coulter Inc. recalled 4 products on 2025-09-24

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2025-09-24
Critical
0
Units

Why these products were recalled

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2025-09-24

    DxH 900 hematology analyzers recalled over erroneously high hemoglobin results

    UniCel DxH 900 Coulter Cellular Analysis Systems configured with the LED HGB photometer are recalled because they may produce erroneously high hemoglobin results on samples with elevated white blood cell counts.

    Product
    UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478
    Category
    Distribution
    37 states
  • HighFDA (Devices)··2025-09-24

    DxH 800 hematology analyzers recalled over erroneously high hemoglobin results

    UniCel DxH 800 Coulter Cellular Analysis Systems configured with the LED HGB photometer are recalled because they may produce erroneously high hemoglobin results on samples with elevated white blood cell counts.

    Product
    UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322
    Category
    Distribution
    37 states
  • HighFDA (Devices)··2025-09-24

    DxH 600 hematology analyzers recalled over erroneously high hemoglobin results

    UniCel DxH 600 Coulter Cellular Analysis Systems configured with the LED HGB photometer are recalled because they may produce erroneously high hemoglobin results on samples with elevated white blood cell counts.

    Product
    UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858
    Category
    Distribution
    37 states
  • HighFDA (Devices)··2025-09-24

    Hematology analyzers recalled over erroneously high hemoglobin results in some samples

    UniCel DxH 690T Coulter Cellular Analysis Systems configured with the LED HGB photometer are recalled because they may produce erroneously high hemoglobin results on samples with elevated white blood cell counts.

    Product
    UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520
    Category
    Distribution
    37 states