Landauer nanoDot radiation dosimeters recalled for accuracy drift
Landauer is recalling nanoDot D2DNS radiation dosimeters worldwide because some units may read outside the specified ±5.5% accuracy range due to a potential non-conformance in the Optical Stimulated Luminescence material.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall affecting a risk-of-harm product (radiation dosimeter) where accuracy is critical to function. No illnesses, injuries, or hospitalizations have been reported. Per the rubric, risk-of-harm products without reported injury are High (3), not Severe.
Plain-English summary
Landauer is recalling nanoDot D2DNS radiation monitoring dosimeters (Model 03066-OTO), which are used with microSTAR readers. All batch numbers distributed worldwide are affected.
Investigation found that some dosimeters may operate outside the specified accuracy range of ±5.5%. The issue is a potential non-conformance in the Optical Stimulated Luminescence (OSL) material used in the dosimeter. The OSL material is stimulated by beaming after radiation exposure and emits fluorescence proportional to the radiation dose received.
The recalled product
- Product
- QC (80 kVp) screened nanoDot D2DNS, Model Number 03066-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Manufacturer
- Landauer
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI/DI 0860003399903
- all batch numbers
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 38 recalled by Landauer under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 38 products in this recall →Related recalls
Same manufacturer · Landauer
See all →- ModerateLandauer microStar Reader dosimetry system discontinued for medical applications
FDA (Devices) · 2024-03-27
- HighLandauer microSTARii reader recalled for measurement inaccuracy
FDA (Devices) · 2024-01-03
- HighLandauer nanoDot D2DXS Dosimeter Recall for Measurement Accuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot A Dosimeters Recalled for Potential Measurement Inaccuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot Radiation Dosimeters Recalled Due to Potential Accuracy Issues
FDA (Devices) · 2023-09-20
Same hazard · accuracy drift
See all →- HighLandauer nanoDot radiation dosimeters recalled for accuracy drift
FDA (Devices) · 2023-09-20
Similar recalls
- HighLandauer microSTARii reader recalled for measurement inaccuracy
FDA (Devices) · 2024-01-03
- ModerateLandauer microStar Reader dosimetry system discontinued for medical applications
FDA (Devices) · 2024-03-27
- HighPTW UNIDOS webline Dosimeters Recalled for Potential Radiation Over-Exposure
FDA (Devices) · 2015-09-02
- HighAngiodynamics Recalls OARtrac PLUS Clinical Detector Units Due to Accuracy Issues
FDA (Devices) · 2019-12-25
- SevereRayStation 8B Software Recall for Dose Calculation Errors
FDA (Devices) · 2020-06-17
- HighSiemens Linear Accelerators Recalled for Potential Radiation Dosing Error
FDA (Devices) · 2012-08-01