FDA (Devices) recall action 92647

Landauer recalled 38 products on 2023-09-20

The FDA (Devices) publishes a recall like this one as 38 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 38 pages.

Products
38
Announced
2023-09-20
Critical
0
Units
1,678,258

Why these products were recalled

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 38

  • ModerateFDA (Devices)··2023-09-20

    Radiation monitoring dosimeters recalled for potential measurement inaccuracy

    Landauer is recalling QC (Cs-137) nanoDot D2DNS radiation dosimeters worldwide because some units may operate outside the specified accuracy range. The recall affects 10,741 units with all batch numbers.

    Product
    QC (Cs-137) nanoDot D2DNS, Model Numbers: a) 18135-000; b) 18135-KIT; c) 18135-SET; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer nanoDot dosimeters recalled for potential accuracy deviation

    Landauer has recalled nanoDot radiation dosimeters due to potential accuracy deviations in the Optical Stimulated Luminescence material used to measure radiation exposure. Affected units may read outside the specified +/-5.5% accuracy range.

    Product
    Constancy (80 kVp) nanoDot D2DNN, Model Numbers: a) 18150-000 (EMEA number VKITCON002); b) 18150-KIT; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    NanoDots radiation dosimeter recalled for potential measurement accuracy issues

    Landauer is recalling NanoDots radiation monitoring dosimeters because some units may not meet specified accuracy of +/-5.5% due to a defect in the luminescence material.

    Product
    NanoDots, Model Number 04297-000; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation Dosimeter Kit Recalled Due to Potential Accuracy Drift

    Landauer is recalling nanoDot radiation dosimeter kits (16,760 units distributed worldwide) because some units may measure radiation exposure outside specified accuracy tolerances, potentially affecting dose measurement accuracy.

    Product
    nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer nanoDot radiation dosimeter may report inaccurate dose readings

    Landauer is recalling certain nanoDot radiation dosimeters worldwide due to a potential flaw in the optical stimulated luminescence material. Affected devices may provide inaccurate radiation dose readings.

    Product
    QC (Cs-137) nanoDot D2DNS, Model Numbers: a) 04225-000 (EMEA number VINLNAN005); b) 04225-KIT; c) 04225-SET; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation Dosimeter Recall for Potential Measurement Accuracy Issues

    Landauer recalls Calibrate nanoDot radiation dosimeters (Model 03064-OTO) worldwide due to potential inaccuracies. Some units may exceed the ±5.5% accuracy specification, resulting in inaccurate radiation exposure readings.

    Product
    Calibrate (80 kVp) screened nanoDot D2DNS, Model Number 03064-OTO; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer NanoDots radiation dosimeters recalled for potential accuracy defect

    Landauer is recalling NanoDots radiation monitoring dosimeters worldwide due to potential accuracy defects. Some units may fall outside the specified +/-5.5% accuracy range.

    Product
    NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 80kVp - Sale; c) BC30084, description: nanoDot QC Set - 80kVp - No Charge; d) BC30088, description: nanoDot Calibration Set - 80kVp - No Charge…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation monitoring dosimeter recalled for potential measurement inaccuracy

    Landauer is recalling Calibrate nanoDot radiation monitoring dosimeters due to a potential defect in the optical luminescence material that may result in inaccurate radiation dose readings.

    Product
    Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 18120-KIT; c) 18120-SET; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Medical Device: Radiation Dosimeter Units May Have Inaccurate Readings

    Landauer nanoDot D2DXN radiation dosimeters may measure outside the specified +/-5.5% accuracy range due to a potential non-conformance in the optical stimulated luminescence material. Approximately 13,950 units were distributed worldwide.

    Product
    nanoDot D2DXN, Model Numbers: a) 03061-1MO; b) 03061-3MO; c) 03061-OTO; d) 03061-SMO; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation dosimeter recalled for potential measurement accuracy deviations

    Landauer recalled nanoDot D2DNS radiation dosimeters worldwide due to potential inaccuracy in Optical Stimulated Luminescence material. Some units may read outside the specified ±5.5% accuracy range.

    Product
    QC (Cs-137) screened nanoDot D2DNS, Model Number 03067-OTO; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation Monitoring Dosimeter Accuracy Issue Affects nanoDot Devices

    Landauer recalled nanoDot C radiation dosimeters due to potential accuracy errors in the optical stimulated luminescence material. Affected units may provide dosimetry readings outside the specified tolerance range of +/-5.5%.

    Product
    nanoDot C - item 3 of 3 for 03500-000, Model Number 03503-000; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2023-09-20

    Landauer nanoDot radiation dosimeters recalled for potential accuracy deviation

    Landauer is recalling nanoDot D2DNS radiation monitoring dosimeters due to a potential defect in the OSL material. Affected devices may provide radiation exposure readings outside the specified +/-5.5% accuracy range.

    Product
    nanoDot D2DNS, Model Numbers: a) 03056-000 (EMEA number VINLNAN002); b) 03056-1MO; c) 03056-3MO; d) 03056-KIT; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer nanoDot B Radiation Dosimeters Recalled for Measurement Accuracy

    Landauer nanoDot B radiation dosimeters may have measurement inaccuracies outside the specified +/-5.5% range, potentially compromising radiation exposure monitoring. 16,820 units worldwide are affected.

    Product
    nanoDot B - item 2 of 3 for 03500-000, Model Number 03502-000; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer nanoDot radiation dosimeters recalled for accuracy drift

    Landauer is recalling nanoDot D2DNN radiation dosimeters because some units may have measurement accuracy outside their specified +/-5.5% range. Worldwide users should contact the manufacturer to verify device performance.

    Product
    nanoDot D2DNN, Model Numbers: a) 03053-1MO; b) 03053-1WK ; c) 03053-3MO; d) 03053-OTO; e) 03053-SMO; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation dosimeter may report inaccurate exposure readings

    Landauer recalls nanoDot D2DNN radiation dosimeters worldwide due to potential accuracy degradation that may cause incorrect exposure readings.

    Product
    nanoDot D2DNN , Model Numbers: a) 04293-000; b) 04293-KIT; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation monitoring dosimeters recalled for potential measurement inaccuracy

    Landauer is recalling Calibrate nanoDot D2DNS radiation dosimeters worldwide due to a potential material defect that may cause readings to fall outside the specified ±5.5% accuracy range.

    Product
    Calibrate (unexposed) nanoDot D2DNS, Model Numbers: a) 18140-000; b) 18140-KIT; c) 18140-SET; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer Radiation Dosimeter Recalled for Potential Measurement Inaccuracy

    Landauer is recalling 1,880 units of the Constancy nanoDot D2DNN radiation dosimeter due to a potential non-conformance that may cause inaccurate radiation readings. The devices were distributed worldwide.

    Product
    Constancy (Cs-137) nanoDot D2DNN, Model Numbers: a) 18155-000 (EMEA number VKITCON001); b) 18155-KIT; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer nanoDot radiation dosimeter units recalled due to accuracy failure

    Landauer is recalling 973 radiation dosimeter units that may not accurately measure radiation exposure due to a potential defect in the light-emitting material used for dose measurement.

    Product
    Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18121-000; b)18121-KIT; c) 18121-SET; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation monitoring dosimeter units recalled for potential accuracy defect

    Landauer is recalling 147 nanoDot D2DNS radiation dosimeters worldwide due to potential non-conformance in the OSL material that may result in readings falling outside the ±5.5% accuracy specification.

    Product
    nanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation monitoring dosimeter accuracy issue prompts FDA Class II recall

    Landauer is recalling 427 nanoDot radiation monitoring dosimeters that may provide inaccurate radiation exposure readings. Some units may fall outside the specified ±5.5% accuracy range due to a manufacturing issue.

    Product
    QC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18131-000; b) 18131-KIT; c) 18131-SET; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2023-09-20

    Landauer nanoDot Radiation Dosimeter Recall for Accuracy Defect

    Landauer is recalling 211,815 nanoDot D2DNS radiation dosimeters worldwide. Some units may be outside the specified ±5.5% accuracy range due to a defect in the optical material, potentially resulting in inaccurate radiation exposure measurements.

    Product
    nanoDot D2DNS, Model Numbers: a) 18100-000; b) 18100-1MO; c) 18100-2MO; d) 18100-KIT; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation Dosimeter Recall: nanoDot D2DNN Accuracy Outside Specifications

    Landauer is recalling nanoDot D2DNN radiation dosimeters worldwide because some units may have measurement accuracy outside the specified +/-5.5% range. Approximately 36,805 units are affected.

    Product
    nanoDot D2DNN, Model Numbers: a) 18105-000; b) 18105-KIT; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Landauer nanoDot radiation dosimeter recalled for potential measurement inaccuracy

    Landauer is recalling nanoDot D2DNS radiation dosimeters worldwide because some units may measure radiation exposure inaccurately, potentially outside the specified +/-5.5% accuracy range. The issue involves the optical stimulated luminescence material used to detect radiation.

    Product
    Calibrate (Cs-137) screened nanoDot D2DNS, Model Number 03065-OTO; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation monitoring dosimeters recalled for potential measurement inaccuracy

    Landauer is recalling Calibrate nanoDot D2DNS radiation dosimeters because they may give inaccurate readings. Some units may read outside the acceptable accuracy range.

    Product
    Calibrate (Cs-137) nanoDot D2DNS, Model Numbers: a) 04224-000 (EMEA number VINLNAN006); b) 04224-KIT; c) 04224-SET; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-09-20

    Radiation Monitoring Dosimeter Potential Accuracy Error: Worldwide Landauer nanoDot Recall

    Landauer is recalling nanoDot D2DXS radiation monitoring dosimeters worldwide due to a potential non-conformance that could cause some units to measure radiation exposure outside the specified +/-5.5% accuracy range.

    Product
    nanoDot D2DXS, Model Numbers: a) 03062-1MO; b) 03062-3MO; c) 03062-OTO; d) 03062-SMO; radiation monitoring dosimeter used with the microSTAR readers
    Category
    Distribution
    0 states