The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2570-2023·Announced 2023-09-20

Radiation monitoring dosimeters recalled for potential measurement inaccuracy

Landauer is recalling Calibrate nanoDot D2DNS radiation dosimeters worldwide due to a potential material defect that may cause readings to fall outside the specified ±5.5% accuracy range.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: This is a Class II recall of a radiation dosimetry device where potential measurement inaccuracy poses a risk of harm to users through inadequate radiation exposure detection. The source reports no illnesses or injuries; the hazard is theoretical rather than demonstrated, placing this at the high end of precautionary recalls.

Plain-English summary

Landauer is recalling certain Calibrate (unexposed) nanoDot D2DNS radiation monitoring dosimeters used with microSTAR readers. The affected model numbers are 18140-000, 18140-KIT, and 18140-SET. A total of 659 units have been identified in this recall, distributed worldwide.

The recall was initiated following reports and investigation indicating that some dosimeters may fall outside the specified accuracy range of ±5.5%. The issue stems from a potential non-conformance in the Optical Stimulated Luminescence (OSL) material, which is beamed after radiation exposure to measure and emit fluorescence proportional to the radiation dose received. If the OSL material does not meet specifications, the dosimeter may provide inaccurate radiation exposure measurements.

Radiation dosimeters are relied upon to detect and quantify radiation exposure for safety monitoring purposes. Inaccurate measurements could fail to properly warn users of excessive radiation exposure or could mask inadequate protection measures.

Customers in possession of affected units should contact Landauer directly for remediation options, which may include replacement or recalibration.

The recalled product

Product
Calibrate (unexposed) nanoDot D2DNS, Model Numbers: a) 18140-000; b) 18140-KIT; c) 18140-SET; radiation monitoring dosimeter used with the microSTAR readers
Manufacturer
Landauer
Affected units
659

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • UDI/DI 0860003399903
  • all batch numbers

Distribution

Distribution scope not specified by the agency.

Part of a larger recall action

This product is one of 38 recalled by Landauer under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 38 products in this recall →

Similar recalls