The Recall Desk
HighFDA (Devices)·Z-2073-2012·Announced 2012-08-01

Siemens Linear Accelerators Recalled for Potential Radiation Dosing Error

Siemens is recalling 18 ARTISTE and ONCOR linear accelerators due to a rare software issue that could cause radiation beams to be delivered twice.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a high-risk medical device where a software error could lead to unintended radiation exposure (risk-of-harm product), but the source text describes the scenario as 'very rare' and does not report any specific injuries or illnesses, fitting the criteria for a High severity score.

Plain-English summary

Siemens Medical Solutions USA, Inc. is recalling 18 ARTISTE and ONCOR Digital Linear Accelerators equipped with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X. These devices are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

The recall is issued due to a potential safety issue involving the IMMEDIATE RESUMPTION feature. In a very rare scenario, if a Control Console interlock occurs during the delivery of flattening-filter-free (Multiple-X) beams or segments, the system may send the previously completed beam or segment for delivery again. This could result in the repetition of already delivered Monitor Units, leading to an unintended radiation dose.

The affected units were distributed worldwide, including in the United States (specifically Florida), Germany, India, Poland, Thailand, and the Republic of Korea. Affected part numbers include 08139789 (ARTISTE), 05863472 (ONCOR Avant Garde), 05857912 (ONCOR Impression Plus), and 07360717 (ONCOR Expression). Healthcare facilities should contact Siemens Medical Solutions USA, Inc. for further instructions on managing the affected equipment.

The recalled product

Product
ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG. The ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatme
Affected units
18

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (8)

  • ARTISTE
  • Part #: 08139789
  • ONCOR Avant Garde
  • Part #: 05863472
  • ONCOR Impression Plus
  • Part #: 05857912
  • ONCOR Expression
  • Part #: 07360717

Distribution

Distributed in 1 state:

Same manufacturer · Siemens Medical Solutions USA, Inc

See all →