The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2318-2021·Announced 2021-08-25

CT X-ray System Software Miscalculates Contrast Volume with Pounds Unit Setting

Philips 728333 Spectral CT 7500 systems with SynchRight software may incorrectly calculate and recommend increased contrast volume when patient weight is set to pounds. Affected system should be updated to prevent excess contrast dosing.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: This is a risk-of-harm product where no injuries have been reported. The software calculation error could lead to excess contrast administration, which poses potential for contrast-related complications.

Plain-English summary

Philips is recalling a Spectral CT 7500 computed tomography X-ray system (serial number 10019, software version 5.0.0 SynchRight Option) due to a software calculation error. When the patient weight unit preference is set to pounds in the SynchRight P3T (Personalized Patient Protocol) software, the system will recommend an increased contrast volume for the patient.

Excess contrast in CT imaging can increase risk of contrast-related complications. This software error could lead to unintended higher contrast doses if clinical staff do not independently verify the system's recommendations.

The affected system is distributed in the United States across Arizona, Florida, Louisiana, Minnesota, Nebraska, Nevada, New York, and Pennsylvania.

Imaging centers using this system should contact Philips North America for a software update. Until the correction is applied, clinical staff should independently verify all contrast volume recommendations, particularly when patient weight is set to pounds.

The recalled product

Product
728333 Spectral CT 7500 -Computed Tomography X-ray system
Affected units
1

Distribution

Distributed nationwide across the United States.

The notice also names 8 states specifically:

Part of a larger recall action

This product is one of 5 recalled by Philips North America Llc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 5 products in this recall →

Same manufacturer · Philips North America Llc

See all →

Same hazard · dosage miscalculation

See all →