Philips MR 7700 MRI Systems Recalled for Software Issue
Philips is recalling 37 MR 7700 MRI systems due to a software issue that can terminate examinations during image reconstruction.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any injuries, illnesses, or hospitalizations. The hazard is a software malfunction that terminates examinations, which fits the criteria for a moderate severity recall involving a risk of harm where injury has not yet been reported or is theoretical.
Plain-English summary
Philips North America LLC is recalling 37 units of the MR 7700 and Upgrade to MR 7700 MRI systems. The affected model numbers are 782120, 782153, and 782130. These devices were distributed worldwide, including in the United States and numerous international locations.
The recall is due to a software issue that will terminate an examination during the image reconstruction process if scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. This issue affects the specific serial numbers listed in the recall documentation.
Healthcare facilities should contact Philips North America LLC for instructions on how to resolve the software issue. Patients and consumers should consult with their healthcare provider regarding any examinations performed on these devices.
The recalled product
- Product
- MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);
- Manufacturer
- Philips North America Llc
- Category
- Medical Device — MRI Systems
- Affected units
- 37
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1. Model Number (REF): 782120
- UDI-DI: 884838104112
- Serial Numbers: 65002
- 65003
- 65014
- 65015
- 65050
- 65029
- 65056
- 65018
- 65021
- 65035
- 65038
- 65000
- 65025
- 65006
- 65008
- 65043
- 2. Model Number (REF): 782153
- UDI-DI: 884838112858
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Philips North America Llc
See all →- HighPhilips Spectral CT 7500 Software Recall for Radiation Exposure Risk
FDA (Devices) · 2026-10-07
- ModeratePhilips MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModeratePhilips Ingenia MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModeratePhilips MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModeratePhilips Intera 1.5T MRI Systems Recalled for Software Issue
FDA (Devices) · 2026-09-30
Same hazard · software defect
See all →- SevereRam Recalls 2022-2024 1500 Classic Trucks for ABS Software Defect
NHTSA · 2026-11-19
- SevereAbiomed Recalls 428 Automated Impella Controllers for Hardware and Software Defects
FDA (Devices) · 2026-11-04
- SevereAbiomed Recalls Impella Optical Controllers Due to Software and Hardware Defects
FDA (Devices) · 2026-11-04
- SevereAbiomed Recalls Impella Connect Optical AIC Units for Hardware and Software Defects
FDA (Devices) · 2026-11-04
- ModerateMedtronic CareLink SmartSync RF Head Patient Connector Software Recall
FDA (Devices) · 2026-11-04
Similar recalls
- ModeratePhilips MR 5300 MRI Scanner Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModeratePhilips MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModeratePhilips Upgrade to MR 7700 Medical Device Alignment Error Recall
FDA (Devices) · 2025-05-28
- ModeratePhilips Achieva XR MRI Systems Recalled for Software Memory Issue
FDA (Devices) · 2026-09-30
- ModeratePhilips MR 7700 MRE Upgrade MRI Equipment Stiffness Value Error Risk
FDA (Devices) · 2026-05-06
- ModeratePhilips Ingenia MRI Software Recall for Image Reconstruction Failure