The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-3242-2026·Announced 2026-09-30

Philips MR 7700 MRI Systems Recalled for Software Issue

Philips is recalling 37 MR 7700 MRI systems due to a software issue that can terminate examinations during image reconstruction.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any injuries, illnesses, or hospitalizations. The hazard is a software malfunction that terminates examinations, which fits the criteria for a moderate severity recall involving a risk of harm where injury has not yet been reported or is theoretical.

Plain-English summary

Philips North America LLC is recalling 37 units of the MR 7700 and Upgrade to MR 7700 MRI systems. The affected model numbers are 782120, 782153, and 782130. These devices were distributed worldwide, including in the United States and numerous international locations.

The recall is due to a software issue that will terminate an examination during the image reconstruction process if scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. This issue affects the specific serial numbers listed in the recall documentation.

Healthcare facilities should contact Philips North America LLC for instructions on how to resolve the software issue. Patients and consumers should consult with their healthcare provider regarding any examinations performed on these devices.

The recalled product

Product
MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);
Affected units
37

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • 1. Model Number (REF): 782120
  • UDI-DI: 884838104112
  • Serial Numbers: 65002
  • 65003
  • 65014
  • 65015
  • 65050
  • 65029
  • 65056
  • 65018
  • 65021
  • 65035
  • 65038
  • 65000
  • 65025
  • 65006
  • 65008
  • 65043
  • 2. Model Number (REF): 782153
  • UDI-DI: 884838112858

Distribution

Distributed nationwide across the United States.

Same manufacturer · Philips North America Llc

See all →