The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-3241-2026·Announced 2026-09-30

Philips MR 5300 MRI Scanner Software Recall for Image Reconstruction Failure

Philips is recalling 611 MR 5300 MRI scanners due to a software issue that can terminate examinations during image reconstruction if system memory is exceeded.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, which typically corresponds to a moderate severity where the hazard is unlikely to cause serious adverse health consequences but could cause temporary or medically reversible harm.

Plain-English summary

Philips North America LLC is recalling 611 units of the MR 5300 MRI scanner (Model Numbers 782110 and 782152). The recall is due to a software issue that can cause the system to terminate an examination during the image reconstruction process. This failure occurs if scan parameters are manually adjusted in a way that causes reconstruction requirements to exceed the available system memory.

The affected devices were distributed worldwide, including in the United States and numerous other countries. The recall is classified as Class II by the FDA, indicating that use of the product could, in some cases, cause temporary or medically reversible health consequences.

Healthcare facilities should contact Philips North America LLC for instructions on how to resolve the software issue. Patients who have undergone scans on these devices should consult their healthcare provider if they have concerns about the integrity of their imaging results.

The recalled product

Product
MR 5300 Model Number (REF): 1. 782110, 2. 782152;
Affected units
611

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • 1. Model Number (REF): 782110
  • UDI-DI: 884838099364
  • Serial Numbers: 63383
  • 63384
  • 63385
  • 63386
  • 63422
  • 63435
  • 63643
  • 63346
  • 63021
  • 63054
  • 63091
  • 63092
  • 63113
  • 63313
  • 63314
  • 63315
  • 63322
  • 63521

Distribution

Distributed nationwide across the United States.

Same manufacturer · Philips North America Llc

See all →