The Recall Desk

FDA (Devices) recall action 88336

Philips North America Llc recalled 5 products on 2021-08-25

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2021-08-25
Critical
0
Units
9

Why these products were recalled

When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–5 of 5

  • HighFDA (Devices)··2021-08-25

    CT X-ray System Software Miscalculates Contrast Volume with Pounds Unit Setting

    Philips 728333 Spectral CT 7500 systems with SynchRight software may incorrectly calculate and recommend increased contrast volume when patient weight is set to pounds. Affected system should be updated to prevent excess contrast dosing.

    Product
    728333 Spectral CT 7500 -Computed Tomography X-ray system
    Category
    Distribution
    8 states
  • HighFDA (Devices)··2021-08-25

    IQon Spectral CT System Software Error Increases Contrast Volume

    Philips IQon Spectral CT systems can recommend excessive contrast volume when patient weight is set to pounds due to a SynchRight software calculation error, potentially resulting in patient overdose.

    Product
    728332 IQon Spectral CT-Computed Tomography X-ray system
    Category
    Distribution
    8 states
  • HighFDA (Devices)··2021-08-25

    Brilliance iCT Computed Tomography Systems Recalled for Contrast Dosing Error

    Philips Brilliance iCT CT systems with SynchRight P3T software may recommend incorrect contrast volume when patient weight is entered in pounds, potentially affecting patient safety.

    Product
    728306 Brilliance iCT-Computed Tomography X-ray systems 728305 Brilliance iCT Upgrades
    Category
    Distribution
    8 states
  • HighFDA (Devices)··2021-08-25

    CT5000 X-ray System Contrast Dose Calculation Error Under Pounds Setting

    Philips CT5000 X-ray systems may incorrectly calculate contrast volume doses when patient weight units are set to pounds. The software error could recommend higher-than-intended contrast volumes.

    Product
    Model 728321/728323 CT5000 Ingenuity Plus/Pro/Premium-Computed Tomography X-ray system
    Category
    Distribution
    8 states
  • HighFDA (Devices)··2021-08-25

    CT system software error recommends excessive contrast volume

    A software defect in Philips CT imaging systems causes incorrect contrast volume calculations when patient weight units are set to pounds, resulting in increased contrast dosing recommendations. Three units with specific serial numbers are affected.

    Product
    728327 Ingenuity CT Upgrades-Computed Tomography X-ray system
    Category
    Distribution
    8 states