Automated diagnostic analyzer from Abbott recalled for waste pressure buildup
Abbott is recalling 260 Alinity s System analyzers due to a design defect allowing waste pressure to build up and spray operators. No injuries reported.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II without reported illnesses or injuries. The design defect poses a potential spray hazard to operators, but no injuries have been reported. Per the rubric, risk-of-harm products without reported injury are classified as High.
Plain-English summary
Abbott Laboratories, Inc. is recalling the Alinity s System (Model LN 06P16-01, Serial Numbers AS1001 through AS1330), an automated immunoassay analyzer used in blood donor screening, plasma screening, and organ donor centers, hospitals, and reference laboratories. A total of 260 systems have been distributed worldwide.
A design defect in both the hardware and software can allow liquid waste pressure to build up within the system. This pressure buildup has the potential to cause waste material to spray toward users and operators during normal operation.
The recalled systems are distributed worldwide, including in the United States, Australia, and numerous other countries in Europe, the Middle East, Asia, and Africa. Affected facilities include blood donor centers, organ donor centers, hospitals, and reference laboratories.
The recalled product
- Product
- Alinity s System REF 06P16-01. An automated immunoassay analyzer for In Vitro diagnostic use in blood donor screening, plasma and plasmapheresis screening, and organ donor centers, hospitals, and reference laboratories.
- Manufacturer
- Abbott Laboratories, Inc
- Affected units
- 260
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Model Number: LN 06P16-01 Serial numbers AS1001 through AS1330
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Abbott Laboratories, Inc
See all →- HighAbbott Alinity s System software error affects diagnostic analyzer wash cycle
FDA (Devices) · 2022-03-23
- HighAbbott ARCHITECT C System Mixer Recall Due to Blade Separation Risk
FDA (Devices) · 2020-04-15
- ModerateAbbott APS Accelerator Modules Recalled for Missing Pacemaker Safety Label
FDA (Devices) · 2019-10-02
- ModerateAbbott Transferrin Assay Recalled for Instruction Manual Updates
FDA (Devices) · 2019-07-31
- ModerateAbbott Immunoglobulin G Diagnostic Assay Recall for IFU Updates
FDA (Devices) · 2019-07-31
Same hazard · pressure buildup
See all →- Severe
- CriticalMRI upgrade recalled over ice blockage in helium venting paths
FDA (Devices) · 2025-11-19
- CriticalMAGNETOM Verio Dot systems recalled over magnet venting ice blockage risk
FDA (Devices) · 2025-10-08
- CriticalMAGNETOM Cima.X systems recalled over ice blockage in helium venting system
FDA (Devices) · 2025-10-08
- CriticalMAGNETOM Vida Fit MRI systems recalled over magnet vent ice blockage
FDA (Devices) · 2025-10-08
Similar recalls
- HighAbbott Alinity s System software error affects diagnostic analyzer wash cycle
FDA (Devices) · 2022-03-23
- ModerateAbbott Alinity Reaction Vessels Recalled for Potential Contaminant
FDA (Devices) · 2023-01-25
- HighAlinity hq Analyzer May Generate Incorrect Blood Test Results
FDA (Devices) · 2024-08-28
- SevereAbbott Alinity m Resp-4-Plex Respiratory Diagnostic Kit Recalled for False Positive Results
FDA (Devices) · 2021-10-20
- HighDiagnostic System Software Defects Affecting Sample Processing and Test Accuracy
FDA (Devices) · 2022-11-30
- ModerateAbbott Recalls Alinity i Processing Modules Due to Gear Pump Assembly Issue