MAGNETOM Vida Fit MRI systems recalled over magnet vent ice blockage

Siemens is recalling 41 MAGNETOM Vida Fit MRI systems because an ice blockage in the magnet venting system could stop helium escaping during a quench.

What this means for you

Highest-severity recall. Risk of serious injury or death is documented or strongly suspected. Stop using the product now and follow the agency's instructions.

Our severity reasoning: Scored 5 because the FDA classified this recall as Class I, which the rubric assigns the Critical tier.

Plain-English summary

MAGNETOM Vida Fit MRI systems, model number 11410481, are being recalled because an ice blockage may form, or may already exist, within the magnet venting system.

In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. That pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

The recall covers 41 units, distributed worldwide including nationwide in the United States. The FDA has classified the recall as Class I.

Operating sites should follow the manufacturer's instructions for inspection and correction of the venting system.

The recalled product

Product
MAGNETOM Vida Fit. Model Number: 11410481.
Manufacturer
Siemens Medical Solutions USA, Inc
Hazard
  • helium-leak
  • pressure-buildup
  • venting-failure
  • quench

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Model Number: 11410481. UDI Numbers: (01)04056869245188(21)202129
  • (01)04056869245188(21)202025
  • (01)04056869245188(21)202139
  • (01)04056869245188(21)202016
  • (01)04056869245188(21)202051
  • (01)04056869245188(21)202055
  • (01)04056869245188(21)202114
  • (01)04056869245188(21)202119
  • (01)04056869245188(21)202013
  • (01)04056869245188(21)202050
  • (01)04056869245188(21)202102
  • (01)04056869245188(21)202160
  • (01)04056869245188(21)202127
  • (01)04056869245188(21)202071
  • (01)04056869245188(21)202078
  • (01)04056869245188(21)202123
  • (01)04056869245188(21)202082
  • (01)04056869245188(21)202107
  • (01)04056869245188(21)202031
  • (01)04056869245188(21)202110

Distribution

Distributed nationwide across the United States.